Literature DB >> 24754296

Technical challenges encountered during deployment of the ovation abdominal aortic stent-graft system.

Efstratios Georgakarakos1, George Trellopoulos, Chris V Ioannou, Dimitrios Tsetis.   

Abstract

PURPOSE: To describe technical challenges encountered using the Ovation endograft for abdominal aortic aneurysms and suggest tips and maneuvers for successful management. TECHNIQUE: Deployment of the Ovation's unsupported main body is often associated with the anteroposterior arrangement of the limb gates instead of the usual side-by-side configuration, rendering contralateral catheterization challenging. Successful catheterization of the contralateral limb can be confirmed by lateral deviation of the ipsilateral stiff guidewire caused by balloon inflation within the contralateral limb. Moreover, failure to cannulate the contralateral limb gate due to persistent impingement of the guidewire or compromise of the inflated rings can be overcome using the transbrachial antegrade approach.
CONCLUSION: Awareness of certain aspects associated with the use of the Ovation device and adoption of the adjunctive techniques herein described helps the interventionist avoid unexpected challenges or manage technical difficulties to maximize the device's safety and effectiveness.

Entities:  

Mesh:

Year:  2014        PMID: 24754296     DOI: 10.1583/13-4515MR.1

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  3 in total

1.  Endovascular aneurysm repair with the Ovation TriVascular Stent Graft System utilizing a predominantly percutaneous approach under local anaesthesia.

Authors:  C V Ioannou; N Kontopodis; E Kehagias; A Papaioannou; A Kafetzakis; G Papadopoulos; D Pantidis; D Tsetis
Journal:  Br J Radiol       Date:  2015-05-12       Impact factor: 3.039

2.  Preliminary experience with the use of ultra-low profile endografts.

Authors:  Daniela Mazzaccaro; Giovanni Malacrida; Bruno Amato; Salvatore Alessio Angileri; Anna Maria Ierardi; Giovanni Nano
Journal:  Diagn Interv Radiol       Date:  2017 Nov-Dec       Impact factor: 2.630

3.  A Misdiagnosed Case of Endoleak Type-III Caused by Inadvertent Bilateral Limb Insertion in the Same Limb Gate of the Ovation Endograft.

Authors:  Efstratios Georgakarakos; Andreas Koutsoumpelis; Panagiotis Kostoglou; Kalliopi-Maria Tasopoulou; Christos Argyriou
Journal:  Aorta (Stamford)       Date:  2022-05-31
  3 in total