Intraoperative measurement for a new transcutaneous bone conduction hearing implant.

Johannes Schnabl1, Astrid Wolf-Magele, Stefan Marcel Pok, Philipp Schoerg, Lena Hirtler, Max Schloegel, Georg Sprinzl.   

Abstract

OBJECTIVE: To investigate the possibility of using a modified reverse transfer function (RTF) measurement intraoperatively during surgery of a new transcutaneous bone conduction hearing implant to evaluate the status of the device.
METHODS: Tests were performed on a cadaver skull (preclinically) and two conductive hearing loss patients implanted with a new transcutaneous bone conduction implant. During intraoperative activation, the RTF was measured using a microphone attached perpendicularly and directly to the skin in the middle section of the forehead.
RESULTS: The RTF could be measured for all frequencies from 500 to 6, 000 Hz.
CONCLUSION: The usage of an intraoperative RTF measurement may be a good method to verify the mechanical coupling of the bone conduction floating mass transducer and to test the functional integrity of the implant in an objective way.

Entities:  

Mesh:

Year:  2014        PMID: 24751748     DOI: 10.1097/MAO.0000000000000351

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  5 in total

1.  Three-Year Follow-Up with the Bone Conduction Implant.

Authors:  Ann-Charlotte Persson; Sabine Reinfeldt; Bo Håkansson; Cristina Rigato; Karl-Johan Fredén Jansson; Måns Eeg-Olofsson
Journal:  Audiol Neurootol       Date:  2020-04-08       Impact factor: 1.854

2.  A Bone Conduction Implantable Device as a Functional Treatment Option in Unilateral Microtia with Bilateral Stapes Ankylosis: A Report of Two Cases.

Authors:  Diego Zanetti; Federica Di Berardino
Journal:  Am J Case Rep       Date:  2018-01-23

Review 3.  Systematic review of outcome domains and instruments used in designs of clinical trials for interventions that seek to restore bilateral and binaural hearing in adults with unilateral severe to profound sensorineural hearing loss ('single-sided deafness').

Authors:  Roulla Katiri; Deborah A Hall; Catherine F Killan; Sandra Smith; Pattarawadee Prayuenyong; Pádraig T Kitterick
Journal:  Trials       Date:  2021-03-20       Impact factor: 2.279

4.  Active transcutaneous bone conduction hearing implants: Systematic review and meta-analysis.

Authors:  Astrid Magele; Philipp Schoerg; Barbara Stanek; Bernhard Gradl; Georg Mathias Sprinzl
Journal:  PLoS One       Date:  2019-09-16       Impact factor: 3.240

5.  A Precision Driver Device for Intraoperative Stimulation of a Bone Conduction Implant.

Authors:  Mohammad Ghoncheh; Thomas Lenarz; Hannes Maier
Journal:  Sci Rep       Date:  2020-02-04       Impact factor: 4.379