Abstract

OBJECTIVES: The aim of this investigation was to perform a biomechanical comparison between 1- and 2-screw systems used for the treatment of intertrochanteric fractures for centralized and decentralized placement of femoral neck screws of failure loads, stiffness, survival rates, tip apex distance (TAD), and failure mode.
METHODS: As fracture model, an AO 31A2.3 fracture was used. Twelve pairs of human cadaver femora were tested. Femoral neck screws were implanted in the femoral head in center/center, posterior/central, and anterior/superior position in axial/frontal plane. A single-screw system (Gamma 3 Locking Nail; Stryker GmbH & Co. KG) and a 2-screw system (Trigen-Intertan; Smith & Nephew GmbH) were used. To simulate the load in situ, a cyclic load was carried for 10,000 cycles in a material testing machine. If no cyclic failure occurred, femora were loaded until the failure. The systems were compared according to the stiffness, survivability through 10 k cycles, TAD, and load to failure.
RESULTS: None of the tested bones failed at center/center location in the decentralized positions 3 Gamma Nail and 2 Intertan specimens failed during cyclic testing. The 2-screw system resisted higher forces in all positions (Gamma: 5370N ± 1924, Intertan: 7650N ± 2043; P = 0.014).
CONCLUSIONS: Based on these data, it is clear that both the nail systems showed a higher biomechanical stability with a lower TAD. The 2 specimens that failed with the Intertan in the cyclic tests had a TAD ≥49 mm. The cutout failures that we detected during cyclic testing in the Gamma system had a TAD ≥30 mm. Thus, it is clear that the TAD affects failure independent of the implant used. With a less than ideal lag screw placement, however, the Intertan system with 2 integrated screws was able to withstand higher loads in this study.

Entities:  

Mesh:

Year:  2014        PMID: 24751606     DOI: 10.1097/BOT.0000000000000008

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  17 in total

Review 1.  [Pertrochanteric femoral fractures in the elderly].

Authors:  G H Sandmann; P Biberthaler
Journal:  Unfallchirurg       Date:  2015-05       Impact factor: 1.000

Review 2.  Is rotation the mode of failure in pertrochanteric fractures fixed with nails? Theoretical approach and illustrative cases.

Authors:  C Kokoroghiannis; D Vasilakos; K Zisis; G Dimitriou; E Pappa; D Evangelopoulos
Journal:  Eur J Orthop Surg Traumatol       Date:  2019-09-20

3.  Comparison of Second and Third-Generation Nails in the Treatment of Intertrochanteric Fracture: Screws versus Helical Blades.

Authors:  Mert Kumbaraci; Levent Karapinar; Ali Turgut
Journal:  Eurasian J Med       Date:  2017-02

4.  How evolution of the nailing system improves results and reduces orthopedic complications: more than 2000 cases of trochanteric fractures treated with the Gamma Nail System.

Authors:  R Pascarella; R Fantasia; A Maresca; C Bettuzzi; L Amendola; S Violini; F Cuoghi; P Sangiovanni; S Cerbasi; S Boriani; D S Tigani
Journal:  Musculoskelet Surg       Date:  2015-12-14

5.  InterTan nail versus Gamma3 nail for intramedullary nailing of unstable trochanteric fractures.

Authors:  Dankai Wu; Guangkai Ren; Chuangang Peng; Xuanlin Zheng; Fengmin Mao; Yueyang Zhang
Journal:  Diagn Pathol       Date:  2014-10-01       Impact factor: 2.644

6.  Impact of tip-apex distance and femoral head lag screw position on treatment outcomes of unstable intertrochanteric fractures using cephalomedullary nails.

Authors:  Cheng-Hung Lee; Kuo-Chih Su; Kun-Hui Chen; Chien-Chou Pan; Yun-Che Wu
Journal:  J Int Med Res       Date:  2018-05-30       Impact factor: 1.671

7.  Comparison of clinical outcomes with InterTan vs Gamma nail or PFNA in the treatment of intertrochanteric fractures: A meta-analysis.

Authors:  Jian-Xiong Ma; Ming-Jie Kuang; Zheng-Rui Fan; Fei Xing; Yun-Long Zhao; Lu-Kai Zhang; Heng-Ting Chen; Chao Han; Xin-Long Ma
Journal:  Sci Rep       Date:  2017-11-21       Impact factor: 4.379

8.  Load distribution between cephalic screws in a dual lag screw trochanteric nail.

Authors:  Julia Henschel; Sebastian Eberle; Peter Augat
Journal:  J Orthop Surg Res       Date:  2016-04-01       Impact factor: 2.359

9.  A retrospective analysis of the InterTan nail and proximal femoral nail anti-rotation-Asia in the treatment of unstable intertrochanteric femur fractures in the elderly.

Authors:  Weiguang Yu; Xinchao Zhang; Xingfei Zhu; Jun Hu; Yunjiang Liu
Journal:  J Orthop Surg Res       Date:  2016-01-15       Impact factor: 2.359

10.  Greater rate of cephalic screw mobilisation following proximal femoral nailing in hip fractures with a tip-apex distance (TAD) and a calcar referenced TAD greater than 25 mm.

Authors:  Rocco Aicale; Nicola Maffulli
Journal:  J Orthop Surg Res       Date:  2018-05-02       Impact factor: 2.359

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.