Literature DB >> 24729560

International intercomparison of measuring instruments for radon/thoron gas and radon short-lived daughter products in the NRPI Prague.

K Jílek1, M Hýža2, L Kotík2, J Thomas2, L Tomášek2.   

Abstract

During the 7th European Conference on Protection Against Radon at Home and at Work held in the autumn of 2013 in Prague, the second intercomparison of measuring instruments for radon and its short-lived decay products and the first intercomparison of radon/thoron gas discriminative passive detectors in mix field of radon/thoron were organised by and held at the Natural Radiation Division of the National Radiation Protection Institute (NRPI) in Prague. In total, 14 laboratories from 11 different countries took part in the 2013 NRPI intercomparison. They submitted both continuous monitors for the measurement of radon gas and equivalent equilibrium radon concentration in a big NRPI chamber (48 m3) and sets of passive detectors including radon/thoron discriminative for the measurement of radon gas in the big chamber and thoron gas in a small thoron chamber (150 dm3).
© The Author 2014. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Mesh:

Substances:

Year:  2014        PMID: 24729560     DOI: 10.1093/rpd/ncu079

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  1 in total

1.  Impact of Wind Speed on Response of Diffusion-Type Radon-Thoron Detectors to Thoron.

Authors:  Yasutaka Omori; Yuki Tamakuma; Eka Djatnika Nugraha; Takahito Suzuki; Miki Arian Saputra; Masahiro Hosoda; Shinji Tokonami
Journal:  Int J Environ Res Public Health       Date:  2020-05-02       Impact factor: 3.390