Evaluation of the automated coagulation analyzer CS-5100 and its utility in high throughput laboratories.

Franz Ratzinger, Klaus G Schmetterer, Helmuth Haslacher, Thomas Perkmann, Sabine Belik, Peter Quehenberger.   

Abstract

BACKGROUND: Automated analyzers are an important component of modern laboratories. As a representative of the newest generation of coagulation analyzers, the CS-5100 features several technical refinements including a pre-analytical assessment unit as well as multi-wavelength optical detection units. Therefore, the CS-5100 is supposed to rapidly and accurately perform a broad panel of coagulation tests. In the current study, the CS-5100 was evaluated regarding its precision and practicability in a clinical laboratory setting.
METHODS: The CS-5100 was evaluated regarding its intra- and inter-assay precision using commercially available control samples. RESULTS of patient samples, including hemolytic, icteric and lipemic specimens, measured on the CS-5100 were compared to reference analyzers, which are used in our accredited laboratory.
RESULTS: The coefficients of variation, assessed in the intra- and inter-assay precision analyses were below 5% representatively for most parameters. RESULTS, obtained by the CS-5100 showed predominantly a high comparability to used reference analyzers, with correlation coefficients ranging from 0.857 to 0.990. Only minor ranged systemic or proportional differences were found in Passing-Bablok regression between the CS-5100 and reference analyzers regarding most of the tested parameters. Lipemic samples had a tendency to deteriorate correlation coefficients, but an overall effect of the sample's triglyceride level could be ruled out. In a routine setting, the analyzer reached a sample throughput rate of 160 tests per hour.
CONCLUSIONS: The CS-5100 is able to rapidly and precisely measure patient samples. No considerable influence on test comparability was found for elevated levels of free hemoglobin, bilirubin or triglycerides.

Entities:  

Mesh:

Year:  2014        PMID: 24721721     DOI: 10.1515/cclm-2013-1094

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  2 in total

1.  A comparative evaluation of the CN-6000 haemostasis analyser using coagulation, amidolytic, immuno-turbidometric and light transmission aggregometry assays.

Authors:  Chris Gardiner; Philip Lane; Katy Langley; Hitesh Tailor; Samuel J Machin; Ian J Mackie
Journal:  Int J Lab Hematol       Date:  2020-06-17       Impact factor: 3.450

2.  The Sysmex CS-5100 coagulation analyzer offers comparable analytical performance and excellent throughput capabilities.

Authors:  T Flieder; T Gripp; C Knabbe; I Birschmann
Journal:  Pract Lab Med       Date:  2016-09-21
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.