Abstract
The first HILIC-tandem mass spectrometry (MS/MS) method for determination of vindesine (VDS) in human plasma using vinorelbine as an internal standard (IS) has been developed and validated. Plasma samples clean-up consisted of solid phase extraction with a strata™-X column. The compounds were separated on a HILIC column with an isocratic mobile phase consisting of acetonitrile and 15mM ammonium acetate buffer containing 0.15% formic acid (80:20, v/v). The detection was performed on a triple quadrupole tandem mass spectrometer via electrospray positive ionization (ESI(+)). The ion transitions recorded in multiple reaction monitoring mode were m/z 754.6→123.8 for VDS and 779.4→323.3 for IS, respectively. Linear calibration curves were obtained in the concentration range of 0.3-28ng/ml and the lower limit of quantification for VDS was 0.3ng/ml. The coefficient of variation of the assay precision was less than 13%, and the accuracy exceeded 96%. The developed assay method was successfully applied for the evaluation of population pharmacokinetics of VDS after intravenous infusion of Xi Ai Ke Vial(®) (3mg of Vindesine Sulfate for Injection) to Chinese Han subjects with hematological malignant disorders.Entities:
Keywords: HILIC; MS/MS; Pharmacokinetics; Vindesine; Vinorelbine
Mesh:
Substances:
Year: 2014 PMID: 24721203 DOI: 10.1016/j.jpba.2014.03.017
Source DB: PubMed Journal: J Pharm Biomed Anal ISSN: 0731-7085 Impact factor: 3.935