Paradoxical signal pattern of mediastinal cysts on T2-weighted MR imaging: phantom and clinical study.

Ken Ueda1, Masahiro Yanagawa2, Takashi Ueguchi3, Yukihisa Satoh2, Misa Kawai2, Tomoko Gyobu2, Hiromitsu Sumikawa2, Osamu Honda2, Noriyuki Tomiyama2.   

Abstract

PURPOSE: To evaluate the intracystic MRI (magnetic resonance imaging) signal intensity of mediastinal cystic masses on T2-weighted images.
MATERIALS AND METHODS: A phantom study was performed to evaluate the signal intensity of a mediastinal cystic mass phantom (rubber balloon containing water) adjacent to a cardiac phantom pulsing at the rate of 60/min. T2-weighted images (sequence, fast spin echo [FSE] and single shot fast spin echo [SSFSE]) were acquired for the mediastinal cystic mass phantom. Further, a clinical study was performed in 33 patients (16 men, 17 women; age range, 19-85 years; mean, 65 years) with thymic cysts or pericardial cysts. In all patients, T2-weighted images (FSE and SSFSE) were acquired. The signal intensity of cystic lesion was evaluated and was compared with that of muscle. A region of interest (ROI) was positioned on the standard MR console, and signal intensity of the cystic mass (cSI), that of the muscle (mSI), and the rate of absolute value of cSI-mSI to standard deviation (SD) of background noise (|cSI-mSI|/SD=CNR [contrast-to-noise ratio]) were measured.
RESULTS: The phantom study demonstrated that the rate phantom-ROI/saline-ROI was higher in SSFSE (0.36) than in FSE (0.19). In clinical cases, the degree of the signal intensity was higher in SSFSE than in FSE. The CNR was significantly higher in SSFSE (mean ± standard deviation, 111.0 ± 47.6) than in FSE (72.8 ± 36.6) (p<0.001, Wilcoxon signed-rank test).
CONCLUSIONS: Anterior mediastinal cysts often show lower signal intensity than the original signal intensity of water on T2-weighted images. SSFSE sequence reduces this paradoxical signal pattern on T2-weighted images, which may otherwise cause misinterpretation when assessing cystic lesions.
Copyright © 2014 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Keywords:  MRI (magnetic resonance imaging); mediastinal cyst fast spin echo (FSE); single shot fast spin echo (SSFSE)

Mesh:

Year:  2014        PMID: 24721003     DOI: 10.1016/j.ejrad.2014.03.004

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  1 in total

1.  Mature mediastinal bronchogenic cyst with left pericardial defect: A case report.

Authors:  Xiao Zhu; Lei Zhang; Zhen Tang; Fu-Bao Xing; Xiong Gao; Wen-Bang Chen
Journal:  World J Clin Cases       Date:  2021-12-26       Impact factor: 1.337

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.