Literature DB >> 24718066

Comparison of 2 lumbar total disc replacements: results of a prospective, randomized, controlled, multicenter Food and Drug Administration trial with 24-month follow-up.

Richard D Guyer1, Kenneth Pettine, Jeffrey S Roh, Thomas A Dimmig, Domagoj Coric, Paul C McAfee, Donna D Ohnmeiss.   

Abstract

STUDY
DESIGN: This was a prospective, randomized, controlled multicenter study with 24-month follow-up.
OBJECTIVE: The purpose of this study was to evaluate the safety and efficacy in a Food and Drug Administration Investigation Device Exemption of a new lumbar total disc replacement (TDR) by comparing it to an earlier TDR approved for sale. SUMMARY OF BACKGROUND DATA: Randomized trials have reported TDR to produce results similar or superior to lumbar fusion. Results for various TDRs seem to be similar, but differences in study design and outcome measures pose challenges in definitively comparing devices. The purpose of this study was to perform a direct comparison of 2 lumbar TDRs in a prospective, randomized trial.
METHODS: TDR was performed in 457 patients from 21 sites (261 patients in the investigational group (Kineflex-L Disc; metal-on-metal design anchored with keels, 204 randomized and 57 nonrandomized training cases), and 196 in the control group (CHARITE artificial disc; metal with polyethylene core with teeth for anchoring; 190 randomized and 6 nonrandomized training cases). All patients were treated nonoperatively for single-level symptomatic disc degeneration for at least 6 months prior to surgery. Perioperative data were collected. Clinical outcome data were collected prospectively, as approved by the Food and Drug Administration, through 24-month follow-up. Primary outcome measures used were the Oswestry Disability Index, visual analogue scales assessing pain, patient satisfaction, and reoperations. Success was defined to be at least 15-point improvements in Oswestry Disability Index scores, no reoperation, and no major adverse events. Radiographical measures included range of motion, disc space height, and assessment for device migration, subsidence, and fusion at the TDR level.
RESULTS: There were no significant differences between the groups when comparing operative time, blood loss, or length of hospital stay. Both groups improved significantly on Oswestry Disability Index and visual analogue scale scores (P < 0.01) with no differences between the groups. Success rates were similar (68.1% investigational vs. 67.4% control). At 24-month follow-up, 94.1% of the investigational group and 91.9% of controls were satisfied with outcome. Reoperation was performed in 10.3% of the investigational group and 8.4% of the control group.
CONCLUSION: This prospective, randomized, controlled study comparing 2 TDRs, the first to the authors' knowledge, found the devices produced very similar clinical outcomes. Both groups improved significantly by 6 weeks postoperatively and remained improved throughout follow-up with a high patient satisfaction rate.

Entities:  

Mesh:

Year:  2014        PMID: 24718066     DOI: 10.1097/BRS.0000000000000319

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  7 in total

1.  Intraosseous basivertebral nerve ablation for the treatment of chronic low back pain: a prospective randomized double-blind sham-controlled multi-center study.

Authors:  Jeffrey S Fischgrund; A Rhyne; J Franke; R Sasso; S Kitchel; H Bae; C Yeung; E Truumees; M Schaufele; P Yuan; P Vajkoczy; M DePalma; D G Anderson; L Thibodeau; B Meyer
Journal:  Eur Spine J       Date:  2018-02-08       Impact factor: 3.134

2.  In vitro analysis of circumferential joint replacement, including bilateral facet joint replacement with lateral lumber disc prosthesis: a parametric investigation of disc sizing.

Authors:  Mark Moldavsky; Pavel Neumann; Noelle Klocke; Mir Hussain; Brandon S Bucklen
Journal:  Eur Spine J       Date:  2016-09-26       Impact factor: 3.134

3.  Lumbar total disc replacement: does it still need further follow-up?

Authors:  Chun-Kun Park
Journal:  J Spine Surg       Date:  2017-09

4.  We Need to Talk about Lumbar Total Disc Replacement.

Authors:  Stephen Beatty
Journal:  Int J Spine Surg       Date:  2018-08-03

5.  ISASS Policy Statement - Lumbar Artificial Disc.

Authors:  Jack Zigler; Rolando Garcia
Journal:  Int J Spine Surg       Date:  2015-03-12

Review 6.  Lumbar total disc arthroplasty: outdated surgery or here to stay procedure? A systematic review of current literature.

Authors:  Matteo Formica; Stefano Divano; Luca Cavagnaro; Marco Basso; Andrea Zanirato; Carlo Formica; Lamberto Felli
Journal:  J Orthop Traumatol       Date:  2017-07-06

Review 7.  Comparison of Lumbar Total Disc Replacement With Surgical Spinal Fusion for the Treatment of Single-Level Degenerative Disc Disease: A Meta-Analysis of 5-Year Outcomes From Randomized Controlled Trials.

Authors:  Jack Zigler; Matthew F Gornet; Nicole Ferko; Chris Cameron; Francine W Schranck; Leena Patel
Journal:  Global Spine J       Date:  2017-11-16
  7 in total