Literature DB >> 24716836

Optic nerve head assessment: comparison of Cirrus optic coherence tomography and Heidelberg Retinal Tomograph 3.

Assaf Kratz1, Ridia Lim, Ivan Goldberg.   

Abstract

BACKGROUND: The purpose of this study was to analyse the relationship between optic nerve head (ONH) parameters measured by spectral domain optical coherence tomography and confocal scanning laser ophthalmoscope.
DESIGN: Prospective, cross-sectional study. Hospital setting. PARTICIPANTS: One hundred seventy-three subjects (85 glaucoma and 88 normal subjects).
METHODS: One eye from each individual was selected randomly for ONH imaging by the spectral domain Cirrus OCT and Heidelberg Retinal Tomograph 3 (HRT3). MAIN OUTCOME MEASURES: Four ONH parameters that are measured by both technologies (average cup-to-disc ratio [CDR], rim area, disc area and cup volume) were analysed and compared for differences, agreement of the categorical classification, diagnostic sensitivities and specificities and the area under the receiver operating characteristic curves (AUC).
RESULTS: ONH parameters, as determined by the two technologies were significantly different but strongly correlated. Proportional bias was demonstrated for all measurements. The agreement of categorical classification was excellent for CDR (κ = 0.94) and good for rim area and cup volume (κ = 0.63 and 0.71, respectively). The highest sensitivities at fixed specificities were achieved by Cirrus OCT. AUCs for CDR, rim area, disc area and cup volume were not significantly different between the two technologies.
CONCLUSIONS: The diagnostic capability of ONH measurements by both technologies is similar. Paired ONH measurements by Cirrus OCT and HRT3 are strongly correlated but significantly different and proportionally biased. The results preclude interchangeable use of the absolute values, but categorical classification of ONH parameters may be interchangeable in clinical practice.
© 2014 Royal Australian and New Zealand College of Ophthalmologists.

Entities:  

Keywords:  Cirrus Optical Coherence Tomography; Heidelberg Retinal Tomograph 3; optic nerve head measurement

Mesh:

Year:  2014        PMID: 24716836     DOI: 10.1111/ceo.12344

Source DB:  PubMed          Journal:  Clin Exp Ophthalmol        ISSN: 1442-6404            Impact factor:   4.207


  4 in total

1.  Evaluating glaucoma damage: emerging imaging technologies.

Authors:  Tigran Kostanyan; Gadi Wollstein; Joel S Schuman
Journal:  Expert Rev Ophthalmol       Date:  2015-02-09

Review 2.  Optic nerve head and fibre layer imaging for diagnosing glaucoma.

Authors:  Manuele Michelessi; Ersilia Lucenteforte; Francesco Oddone; Miriam Brazzelli; Mariacristina Parravano; Sara Franchi; Sueko M Ng; Gianni Virgili
Journal:  Cochrane Database Syst Rev       Date:  2015-11-30

3.  Comparison of Diagnostic Ability Between Wide-Field Swept-Source Optical Coherence Tomography Imaging Maps and Heidelberg Retina Tomograph 3 Optic Nerve Head Assessment to Discriminate Glaucomatous and Non-glaucomatous Eyes.

Authors:  Dimitrios Kourkoutas; George Triantafyllopoulos; Iordanis Georgiou; Aristotelis Karamaounas; Nikolaos Karamaounas; Konstadinos Sotiropulos; Dimitrios Kapralos
Journal:  Cureus       Date:  2022-08-19

4.  Optic nerve head parameters of high-definition optical coherence tomography and Heidelberg retina tomogram in perimetric and preperimetric glaucoma.

Authors:  Viquar Unnisa Begum; Uday Kumar Addepalli; Sirisha Senthil; Chandra Sekhar Garudadri; Harsha Laxmana Rao
Journal:  Indian J Ophthalmol       Date:  2016-04       Impact factor: 1.848