Literature DB >> 24709608

A cubic gamma camera with an active collimator.

Taewoong Lee1, Wonho Lee2.   

Abstract

Mechanical collimation with photon absorption and electronic collimation using Compton scattering are combined to form a cubic gamma camera with an active collimator. The collimator is made active by constructing it with a uniformly redundant array of patterned Bi4Ge3O12 (BGO) scintillators, which not only attenuates incident radiations but also detects scattered radiation, in a gamma-camera consisting of and five planar CsI(Na) scintillators. The entire module forms a cubic structure that generates images on the basis of radiation interactions from every direction. The coverage angle for detecting scattered radiation is 2π with a detection efficiency approximately 17 times higher than previous systems that comprised only one pair of detectors.
Copyright © 2014 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Active collimator; Cubic gamma camera; Electronic collimation; Mechanical collimation

Year:  2014        PMID: 24709608     DOI: 10.1016/j.apradiso.2014.03.020

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  1 in total

1.  High resolution and sensitivity gamma camera with active septa. A first Monte Carlo study.

Authors:  Victor Ilisie; Laura Moliner; Sandra Oliver; Filomeno Sánchez; Antonio J González; Michael Seimetz; Maria J Rodríguez-Álvarez; Jose Maria Benlloch
Journal:  Sci Rep       Date:  2019-12-05       Impact factor: 4.379