Safia Bhimji1, R Michael Meneghini2.
Abstract
Initial implant stability is crucial to cementless knee arthroplasty fixation. The objective of this study was to evaluate the stability of two baseplates with different fixation features: a monoblock porous tantalum baseplate featuring two hexagonal pegs alone, and a modular design featuring a keel with four adjuvant cruciform pegs. A physiologically relevant test method previously described was used to evaluate compression and liftoff of the baseplates during stair descent. The porous tantalum baseplate with dual-hex peg fixation experienced greater rocking motions and liftoff compared to the baseplate with a keel and adjuvant pegs. Liftoff and displacement motion is likely deleterious and may inhibit biological fixation due to the physical separation of the baseplate from the bone.Entities:
Keywords: cementless baseplates; keels; micromotion; pegs; stability; total knee arthroplasty
Mesh:
Substances:
Year: 2014 PMID: 24709524 DOI: 10.1016/j.arth.2014.02.016