Y Doi1, K Takada, O Nakagaki, K Yoshimura, S Ogawa, T Hiroki, K Arakawa.
Abstract
Cross stimulation in a dual chamber pacing system, in which the atrial stimulus intermittently captured the right ventricle, occurred immediately after pacemaker implantation in a 71-year-old man. It was prevented temporarily by reducing the pacing rate so that P wave synchronous ventricular (VDD) pacing resulted and by reducing the output of the atrial circuit from 5 to 4 volts. Cross stimulation disappeared spontaneously 14 days after surgery.Entities:
Mesh:
Year: 1989 PMID: 2470039 DOI: 10.1111/j.1540-8159.1989.tb02702.x