Guidance for accurate and consistent tissue Doppler velocity measurement: comparison of echocardiographic methods using a simple vendor-independent method for local validation.

Niti M Dhutia1, Massoud Zolgharni2, Keith Willson2, Graham Cole2, Alexandra N Nowbar2, David Dawson3, Sayeh Zielke4, Carol Whelan5, Jim Newton4, Jamil Mayet2, Charlotte H Manisty2, Darrel P Francis2.   

Abstract

BACKGROUND: Variability has been described between different echo machines and different modalities when measuring tissue velocities. We assessed the consistency of tissue velocity measurements across different modalities and different manufacturers in an in vitro model and in patients. Furthermore, we present freely available software tools to repeat these evaluations. METHODS AND
RESULTS: We constructed a simple setup to generate reproducible motion and used it to compare velocities measured using three echocardiographic modalities: M-mode, speckle tracking, and tissue Doppler, with a straightforward, non-ultrasound, optical gold standard. In the clinical phase, 25 patients underwent M-mode, speckle tracking, and tissue Doppler measurements of s', e', and a' velocities. In vitro, the M-mode and speckle tracking velocities agreed with optical assessment. Of the three possible tissue Doppler measurement conventions (outer, middle, and inner edge) only the middle agreed with optical assessment (discrepancy -0.20 (95% CI -0.44 to 0.03) cm/s, P = 0.11, outer +5.19 (4.65 to 5.73) cm/s, P < 0.0001, inner -6.26 (-6.87 to -5.65) cm/s, P < 0.0001). A similar pattern occurred across all four studied manufacturers. M-mode was therefore chosen as the in vivo gold standard. Clinical measurements of s' velocities by speckle tracking and the middle line of the tissue Doppler showed concordance with M-mode, while the outer line overestimated significantly (+1.27(0.96 to 1.59) cm/s, P < 0.0001) and the inner line underestimated (-1.82 (-2.11 to -1.52) cm/s, P < 0.0001).
CONCLUSIONS: Echocardiographic velocity measurements can be more consistent than previously suspected. The statistically modal velocity, found at the centre of the spectral pulsed wave tissue Doppler envelope, most closely represents true tissue velocity. This article includes downloadable, vendor-independent software enabling calibration of echocardiographic machines using a simple, inexpensive in vitro setup. Published on behalf of the European Society of Cardiology. All rights reserved.
© The Author 2014. For permissions please email: journals.permissions@oup.com.

Entities:  

Keywords:  Calibration; Echocardiography; Tissue Doppler; Velocity

Mesh:

Year:  2014        PMID: 24699322     DOI: 10.1093/ehjci/jeu040

Source DB:  PubMed          Journal:  Eur Heart J Cardiovasc Imaging        ISSN: 2047-2404            Impact factor:   6.875


  7 in total

Review 1.  The use of echocardiographic indices in defining and assessing right ventricular systolic function in critical care research.

Authors:  Stephen J Huang; Marek Nalos; Louise Smith; Arvind Rajamani; Anthony S McLean
Journal:  Intensive Care Med       Date:  2018-05-22       Impact factor: 17.440

2.  The relationship between cardiac hemodynamics and exercise tolerance in cystic fibrosis.

Authors:  Erik H Van Iterson; Courtney M Wheatley; Sarah E Baker; Wayne J Morgan; Eric M Snyder
Journal:  Heart Lung       Date:  2016-04-01       Impact factor: 2.210

3.  Detecting impaired myocardial relaxation in sepsis with a novel tissue Doppler parameter (septal e'/s').

Authors:  David J Clancy; Michel Slama; Stephen Huang; Timothy Scully; Anthony S McLean; Sam R Orde
Journal:  Crit Care       Date:  2017-07-14       Impact factor: 9.097

4.  Application of updated guidelines on diastolic dysfunction in patients with severe sepsis and septic shock.

Authors:  David J Clancy; Timothy Scully; Michel Slama; Stephen Huang; Anthony S McLean; Sam R Orde
Journal:  Ann Intensive Care       Date:  2017-12-19       Impact factor: 6.925

5.  Mechanical dyssynchrony and diastolic dysfunction are common in LVH: a pilot correlation study using Doppler echocardiography and CZT gated-SPECT MPI.

Authors:  Szu-Ying Tsai; Shan-Ying Wang; Yu-Chien Shiau; Yen-Wen Wu
Journal:  Sci Rep       Date:  2018-03-08       Impact factor: 4.379

6.  Open-source, vendor-independent, automated multi-beat tissue Doppler echocardiography analysis.

Authors:  Niti M Dhutia; Massoud Zolgharni; Michael Mielewczik; Madalina Negoita; Stefania Sacchi; Karikaran Manoharan; Darrel P Francis; Graham D Cole
Journal:  Int J Cardiovasc Imaging       Date:  2017-02-20       Impact factor: 2.357

Review 7.  Tissue Doppler velocity imaging and event timings in neonates: a guide to image acquisition, measurement, interpretation, and reference values.

Authors:  Eirik Nestaas; Ulf Schubert; Willem P de Boode; Afif El-Khuffash
Journal:  Pediatr Res       Date:  2018-07       Impact factor: 3.756

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.