Abstract

GOALS: Automated flowcytometric analysers are used to analyse complete blood count (CBC) in most patients. They differentiate WBC depending on their size, internal granularity, and nuclear content by producing various scattergrams. In malaria, haemozoin pigment-containing cells (parasitized RBC and phagocytic cells) are able to depolarize the laser beam and therefore exhibit abnormal scattergram. PROCEDURE: A study was conducted to evaluate the diagnostic utility and to determine the sensitivity and specificity of flowcytometric haematology analysers [Sysmex XE-2100 (System 1) and Cell-Dyn 3700 (System 2)] in comparison with the conventional microscopy (QBC, AO, Giemsa) and ICT (Immunochromatographic test) for detection of the malaria parasite.
RESULTS: Fifty-eight of 320 samples were found to be positive for malaria by conventional methods. Considering atypical scattergram, System 1 showed the sensitivity and specificity of 68.9% and 90.6%, respectively; whereas System 2 showed a sensitivity of 100% and specificity of 98.8%.
CONCLUSION: System 2 (Cell-Dyn 3700) was found to be a highly sensitive and specific tool compared to System 1 and conventional methods. Hence, it may be preferred for automated detection of malaria in the blood samples of patients with a differential diagnosis of malaria and those who are subjected to CBC analysis.

Entities:  

Keywords:  Quantitative buffy coat assay; flowcytometric analyser; malaria; microscopy; scattergram

Mesh:

Year:  2014        PMID: 24695479

Source DB:  PubMed          Journal:  Ann Clin Lab Sci        ISSN: 0091-7370            Impact factor:   1.256


  3 in total

1.  Reply to "Modern Hematology Analyzers Are Very Useful for Diagnosis of Malaria and, Crucially, May Help Avoid Misdiagnosis".

Authors:  Blaine A Mathison; Bobbi S Pritt
Journal:  J Clin Microbiol       Date:  2017-11       Impact factor: 5.948

2.  Diagnostic accuracy of fluorescence flow-cytometry technology using Sysmex XN-31 for imported malaria in a non-endemic setting.

Authors:  Stéphane Picot; Thomas Perpoint; Christian Chidiac; Alain Sigal; Etienne Javouhey; Yves Gillet; Laurent Jacquin; Marion Douplat; Karim Tazarourte; Laurent Argaud; Martine Wallon; Charline Miossec; Guillaume Bonnot; Anne-Lise Bienvenu
Journal:  Parasite       Date:  2022-05-31       Impact factor: 3.020

3.  Hemozoin Pigment: An Important Tool for Low Parasitemic Malarial Diagnosis.

Authors:  Sarita Mohapatra; Arnab Ghosh; Ruchi Singh; Dhirendra Pratap Singh; Bhawna Sharma; Jyotish Chandra Samantaray; Manorama Deb; Rajni Gaind
Journal:  Korean J Parasitol       Date:  2016-08-31       Impact factor: 1.341

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.