Literature DB >> 24694128

Ion beam tracking using ultrasound motion detection.

M Prall1, R Kaderka1, N Saito1, C Graeff1, C Bert1, M Durante1, K Parodi2, J Schwaab3, C Sarti3, J Jenne3.   

Abstract

PURPOSE: The use of motion mitigation techniques such as tracking and gating in particle therapy requires real-time knowledge of tumor position with millimeter precision. The aim of this phantom-based study was to evaluate the option of diagnostic ultrasound (US) imaging (sonography) as real-time motion detection method for scanned heavy ion beam irradiation of moving targets.
METHODS: For this pilot experiment, a tumor surrogate was moved inside a water bath along two-dimensional trajectories. A rubber ball was used for this purpose. This ball was moved by a robotic arm in two dimensions lateral to the heavy ion beam. Trajectories having a period of 3 s and peak to peak amplitude of 20 mm were used. Square radiation fields of[Formula: see text] were irradiated on radiosensitive films with a 200 MeV/u beam of calcium ions having a FWHM of 6 mm. Pencil beam scanning and beam tracking were employed. The films were attached on the robotic arm and thus moved with the rubber ball. The position of the rubber ball was continuously measured by a US tracking system (Mediri GmbH, Heidelberg) and sent to the GSI therapy control system (TCS). This position was used as tracking vector. Position reconstruction from the US tracking system and data communication introduced a delay leading to a position error of several millimeters. An artificial neural network (ANN) was implemented in the TCS to predict motion from US measurements and thus to compensate for the delay.
RESULTS: Using ANN delay compensation and large motion amplitudes, the authors could produce irradiation patterns with a few percent inhomogeneity and about 1 mm geometrical conformity.
CONCLUSIONS: This pilot experiment suggests that diagnostic US should be further investigated as dose-free, high frame-rate, and model-independent motion detection method for scanning heavy ion beam irradiation of moving targets.
© 2014 American Association of Physicists in Medicine.

Entities:  

Mesh:

Year:  2014        PMID: 24694128     DOI: 10.1118/1.4868459

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  [Ultrasound motion tracking for radiation therapy].

Authors:  J Jenne; J Schwaab
Journal:  Radiologe       Date:  2015-11       Impact factor: 0.635

Review 2.  Particle therapy of moving targets-the strategies for tumour motion monitoring and moving targets irradiation.

Authors:  Tomasz Kubiak
Journal:  Br J Radiol       Date:  2016-07-19       Impact factor: 3.039

Review 3.  Advances in 4D treatment planning for scanned particle beam therapy - report of dedicated workshops.

Authors:  Christoph Bert; Christian Graeff; Marco Riboldi; Simeon Nill; Guido Baroni; Antje-Christin Knopf
Journal:  Technol Cancer Res Treat       Date:  2013-12-17

Review 4.  Management of Motion and Anatomical Variations in Charged Particle Therapy: Past, Present, and Into the Future.

Authors:  Julia M Pakela; Antje Knopf; Lei Dong; Antoni Rucinski; Wei Zou
Journal:  Front Oncol       Date:  2022-03-09       Impact factor: 6.244

5.  First Steps Toward Ultrasound-Based Motion Compensation for Imaging and Therapy: Calibration with an Optical System and 4D PET Imaging.

Authors:  Julia Schwaab; Christopher Kurz; Cristina Sarti; André Bongers; Frédéric Schoenahl; Christoph Bert; Jürgen Debus; Katia Parodi; Jürgen Walter Jenne
Journal:  Front Oncol       Date:  2015-11-30       Impact factor: 6.244

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.