A pilot study of joint stability at the zirconium or titanium abutment/titanium implant interface.

Yeliz Cavusoglu, Kıvanç Akça, Rıza Gürbüz, Murat Cavit Cehreli.   

Abstract

PURPOSE: To compare the interfaces of loaded and unloaded zirconium and titanium abutments with titanium implants using scanning electron microscopy (SEM).
MATERIALS AND METHODS: Zirconium and titanium abutments (n = 5 per group; four test and one control) were torque-tightened into titanium implants secured into metal blocks, and computer-aided design/computer-assisted manufacture-based zirconium oxide copings were fabricated and cemented to the abutments with temporary resin-based cement. Specimens of each restoration were subjected to cyclic axial and lateral loading of 30 N at 2 Hz for 500,000 cycles using a servohydraulic test system; control specimens were left unloaded. Then, the abutment/implant assemblies were embedded in acrylic resin, sectioned longitudinally along the midline, and inspected under SEM with x-ray microanalysis.
RESULTS: Loosening or fracture of the copings and implant components was not observed after dynamic loading in both groups. SEM and x-ray microanalysis revealed unexpected microleakage of acrylic resin at the interface. Acrylic resin in the implants tightened to the titanium abutments was limited to the cervical part, and the components displayed scratched and smashed regions, suggesting slight deformation of the implant neck. Microleakage and pooling of acrylic resin were observed approaching the screw joint in loaded implants tightened to zirconia abutments, and the amount of microleakage was greater than in the unloaded control specimens, which had a larger microgap than the titanium abutment/titanium implant interface. Loaded zirconia abutments were associated with wear, scratches, and, in one sample, chipping.
CONCLUSIONS: Zirconium abutment/titanium implant interface may be susceptible to wear of the abutment coupled with deformation of the implant neck greater than that associated with the conventional titanium abutment/titanium implant interface under dynamic loading.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24683559     DOI: 10.11607/jomi.3116

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  5 in total

1.  Removal torque evaluation of three different abutment screws for single implant restorations after mechanical cyclic loading.

Authors:  T Paepoemsin; P A Reichart; P Chaijareenont; F P Strietzel; P Khongkhunthian
Journal:  Oral Implantol (Rome)       Date:  2016-11-16

Review 2.  Microleakage at the Different Implant Abutment Interface: A Systematic Review.

Authors:  Sunil Kumar Mishra; Ramesh Chowdhary; Shail Kumari
Journal:  J Clin Diagn Res       Date:  2017-06-01

3.  Potential Causes of Titanium Particle and Ion Release in Implant Dentistry: A Systematic Review.

Authors:  Rafael Delgado-Ruiz; Georgios Romanos
Journal:  Int J Mol Sci       Date:  2018-11-13       Impact factor: 5.923

4.  Fracture resistance of different implant supported ceramic abutment/crown systems.

Authors:  Merve Bankoglu Gungor; Secil Karakoca Nemli; Handan Yilmaz; Cemal Aydin
Journal:  Eur Oral Res       Date:  2019-05-01

5.  Effect of Implant Platform Connection and Abutment Material on Removal Torque and Implant Hexagon Plastic Deformation.

Authors:  Paulo Henrique de Souza Pereira; Marina Amaral; Kusai Baroudi; Rafael Pino Vitti; Mohammad Zakaria Nassani; Laís Regiane da Silva-Concílio
Journal:  Eur J Dent       Date:  2019-12-03
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.