Literature DB >> 24682927

The use of a stand-alone interbody fusion cage in subaxial cervical spine trauma: a preliminary report.

Carlo Brembilla1, Luigi Andrea Lanterna1, Paolo Gritti2, Antonio Signorelli1, Francesco Biroli1.   

Abstract

BACKGROUND: Anterior spinal surgery has a predominant role in the treatment of traumatic lesions of the subaxial cervical spine. Plating is considered indispensable to achieve stability but may cause dysphagia, dysphonia, and adjacent-level ossification. Zero-P (Synthes GmbH, Oberdorf, Switzerland), an anchored interdisc spacer, can be used without an associated plate. The present study aimed to test if this new implant would be associated with a low rate of dysphagia and other short-term complications compared with the standard for anterior spinal fusion surgery and would be able to achieve a solid fusion and maintain correct metamere alignment.
MATERIAL AND METHODS: This is a preliminary presentation of a clinical case series of patients with subaxial cervical injuries who underwent anterior interbody fusion. From July 2009 until September 2011, 12 patients were treated with a Zero-P cage. The data for analysis included operating time compared with the standard for spinal fusion surgery with a cage plus plate construct, intraoperative blood loss, clinical and radiographic results, and complications.
RESULTS: In the postoperative period no patient had neurologic worsening. One patient experienced transient dysphonia and moderate dysphagia. All the patients were followed up for a minimum of 6 months (mean: 13 months; range: 6-27 months). Stability and fusion were obtained in all patients together with correct metamere alignment.
CONCLUSION: We presented the preliminary results of a clinical case series. Our results support the initiation of prospective randomized trials with more patients and longer follow-up. Georg Thieme Verlag KG Stuttgart · New York.

Entities:  

Mesh:

Year:  2014        PMID: 24682927     DOI: 10.1055/s-0034-1368092

Source DB:  PubMed          Journal:  J Neurol Surg A Cent Eur Neurosurg        ISSN: 2193-6315            Impact factor:   1.268


  1 in total

1.  Can the Zero-Profile Implant Be Used for Anterior Cervical Discectomy and Fusion in Traumatic Subaxial Disc Injury? A Preliminary, Retrospective Study.

Authors:  Tae Hun Kim; Dae Hyun Kim; Ki Hong Kim; Young Seok Kwak; Sang Gyu Kwak; Man Kyu Choi
Journal:  J Korean Neurosurg Soc       Date:  2018-08-31
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.