Literature DB >> 24679147

The role of post-operative imaging in cochlear implant surgery: a review of 220 adult cases.

Alice Coombs, Philip J Clamp, Susan Armstrong, Philip J Robinson, Daniel Hajioff.   

Abstract

OBJECTIVES: To determine the incidence of abnormal radiological findings after cochlear implantation and their effect on clinical outcomes.
METHODS: Retrospective review of 220 adult cochlear implants. Clinical records and post-operative plain X-rays were reviewed and compared with pre-operative and 6-month post-operative City University of New York (CUNY) speech scores.
RESULTS: There were no cases of extra-cochlear array misplacement. Imaging showed 20 cases of incomplete array insertion (9.2%), 3 cases of kinking of the array (1.4%), 2 cases of tip rollover (0.9%), and 1 case of apparent array fracture (0.5%). Patient management was not altered by abnormal imaging. Patients with abnormal radiological findings had slightly minor improvements (median 39 vs. 56%) in City University of New York (CUNY) speech discrimination scores at 6 months (Mann-Whitney U test, P = 0.043).
CONCLUSION: All abnormalities on post-operative imaging were minor and did not alter patient management. The future role of post-operative imaging is discussed.

Entities:  

Keywords:  Cochlear implants; Electrode array placement; Hearing loss; Post-operative period; Radiography; Speech discrimination tests

Mesh:

Year:  2014        PMID: 24679147     DOI: 10.1179/1754762814Y.0000000071

Source DB:  PubMed          Journal:  Cochlear Implants Int        ISSN: 1467-0100


  5 in total

1.  Incomplete and false tract insertions in cochlear implantation: retrospective review of surgical and auditory outcomes.

Authors:  Ashish Vashishth; Andrea Fulcheri; Maurizio Guida; Antonio Caruso; Mario Sanna
Journal:  Eur Arch Otorhinolaryngol       Date:  2018-03-05       Impact factor: 2.503

2.  Effectiveness of skull X-RAY to determine cochlear implant insertion depth.

Authors:  Vinay Fernandes; Yiqiao Wang; Robert Yeung; Sean Symons; Vincent Lin
Journal:  J Otolaryngol Head Neck Surg       Date:  2018-09-03

3.  Angular Electrode Insertion Depth and Speech Perception in Adults With a Cochlear Implant: A Systematic Review.

Authors:  Floris Heutink; Simone R de Rijk; Berit M Verbist; Wendy J Huinck; Emmanuel A M Mylanus
Journal:  Otol Neurotol       Date:  2019-08       Impact factor: 2.311

4.  Detection of Extracochlear Electrodes in Cochlear Implants with Electric Field Imaging/Transimpedance Measurements: A Human Cadaver Study.

Authors:  Simone R de Rijk; Yu C Tam; Robert P Carlyon; Manohar L Bance
Journal:  Ear Hear       Date:  2020 Sep/Oct       Impact factor: 3.562

5.  Intra-operative radiological diagnosis of a tip roll-over electrode array displacement using fluoroscopy, when electrophysiological testing is normal: the importance of both techniques in cochlear implant surgery.

Authors:  Octavio Garaycochea; Raquel Manrique-Huarte; Manuel Manrique
Journal:  Braz J Otorhinolaryngol       Date:  2017-06-01
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.