Determination of internal target volume using selective phases of a 4-dimensional computed tomography scan.

Junsheng Cao1, Yunfeng Cui2, Colin E Champ2, Haisong Liu2, Ying Xiao2, Maria Werner-Wasik2, Yan Yu2.   

Abstract

PURPOSE: Internal target volume (ITV) is frequently determined by contouring of gross tumor volumes (GTV) on 10 phases of a 4-dimensional computed tomography (4DCT) study set for lung cancer radiotherapy. This study aimed to investigate the possibility of generating ITV by using selective phases of a 4DCT scan. METHODS AND MATERIALS: The 4DCT scans of 20 patients with lung cancer were included in this study. GTVs were contoured on 10 phases in Focal4D (CMS, St Louis, MO). Different ITVs were derived by encompassing volumes of contours from selective phases. ITV10 was the combination of GTVs on all of the 10 phases and served as the gold standard volume. All of the other ITVs were smaller and within ITV10. The ratios of the volumes of these ITVs to ITV10 were calculated and used as a criterion to determine the similarity of different ITVs to ITV10. ITV2 represented the ITV derived by using end-inhalation and end-exhalation (0% + 50%). ITV3E was derived from contouring the 3 phases at end-inhalation, mid-exhalation, and end-exhalation (0% + 20% + 50%). ITV3I was derived from contouring the 3 phases at end-inhalation, mid-inhalation, and end-exhalation (0% + 70% + 50%). ITV4 was derived by contouring the 4 phases at end-inhalation, mid-inhalation, end-exhalation, and mid-exhalation (0% + 20% + 50% + 70%). ITV6E was derived from contouring the 6 consecutive phases during exhalation (0% + 10% + 20% + 30% + 40% + 50%). ITV6I was derived from contouring the 6 consecutive phases during inhalation (50% + 60% + 70% + 80% + 90% + 0%). The volumes of ITVs were calculated and compared.
RESULTS: ITV6I showed excellent agreement with ITV10 (volume ratio ITV6I/ITV10 = 0.975). ITV4 and ITV6E showed good agreement with ITV10 (ITV6E/ITV10 = 0.939, ITV4/ITV10 = 0.944). The volume ratios ITV3I/ITV10 and ITV3E/ITV10 were 0.927 and 0.906, respectively. ITV2 did not agree well with ITV10 (ITV2/ITV10 = 0.888).
CONCLUSIONS: Contouring all phases during inhalation provides a good estimate of the ITV. However, the ITV may be underestimated if only contouring on 2 extreme phases.
Copyright © 2012 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.

Entities:  

Year:  2011        PMID: 24674122     DOI: 10.1016/j.prro.2011.09.004

Source DB:  PubMed          Journal:  Pract Radiat Oncol        ISSN: 1879-8500


  3 in total

1.  Initial clinical experience with a radiation oncology dedicated open 1.0T MR-simulation.

Authors:  Carri K Glide-Hurst; Ning Wen; David Hearshen; Joshua Kim; Milan Pantelic; Bo Zhao; Tina Mancell; Kenneth Levin; Benjamin Movsas; Indrin J Chetty; M Salim Siddiqui
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

2.  Evaluation of two synchronized external surrogates for 4D CT sorting.

Authors:  Carri K Glide-Hurst; Megan Schwenker Smith; Munther Ajlouni; Indrin J Chetty
Journal:  J Appl Clin Med Phys       Date:  2013-11-04       Impact factor: 2.102

3.  Technical Note: Comparison of the internal target volume (ITV) contours and dose calculations on 4DCT, average CBCT, and 4DCBCT imaging for lung stereotactic body radiation therapy (SBRT).

Authors:  Michael Dumas; Eric Laugeman; Parag Sevak; Karen C Snyder; Weihua Mao; Indrin J Chetty; Munther Ajlouni; Ning Wen
Journal:  J Appl Clin Med Phys       Date:  2020-10-12       Impact factor: 2.243

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.