Operating room fire prevention: creating an electrosurgical unit fire safety device.

William C Culp1, Bradly A Kimbrough, Sarah Luna, Aris J Maguddayao.   

Abstract

OBJECTIVE: To reduce the incidence of surgical fires.
BACKGROUND: Operating room fires represent a potentially life-threatening hazard and are triggered by the electrosurgical unit (ESU) pencil. Carbon dioxide is a fire suppressant and is a routinely used medical gas. We hypothesize that a shroud of protective carbon dioxide covering the tip of the ESU pencil displaces oxygen, thereby preventing fire ignition.
METHODS: Using 3-dimensional modeling techniques, a polymer sleeve was created and attached to an ESU pencil. This sleeve was connected to a carbon dioxide source and directed the gas through multiple precisely angled ports, generating a cone of fire-suppressive carbon dioxide surrounding the active pencil tip. This device was evaluated in a flammability test chamber containing 21%, 50%, and 100% oxygen with sustained ESU activation. The sleeve was tested with and without carbon dioxide (control) until a fuel was ignited or 30 seconds elapsed. Time to ignition was measured by high-speed videography.
RESULTS: Fires were ignited with each control trial (15/15 trials). The control group median ± SD ignition time in 21% oxygen was 3.0 ± 2.4 seconds, in 50% oxygen was 0.1 ± 1.8 seconds, and in 100% oxygen was 0.03 ± 0.1 seconds. No fire was observed when the fire safety device was used in all concentrations of oxygen (0/15 trials; P < 0.0001). The exact 95% confidence interval for absolute risk reduction of fire ignition was 76% to 100%.
CONCLUSIONS: A sleeve creating a cone of protective carbon dioxide gas enshrouding the sparks from an ESU pencil effectively prevents fire in a high-flammability model. Clinical application of this device may reduce the incidence of operating room fires.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24670856     DOI: 10.1097/SLA.0000000000000654

Source DB:  PubMed          Journal:  Ann Surg        ISSN: 0003-4932            Impact factor:   12.969


  2 in total

1.  Carbon dioxide can eliminate operating room fires from alcohol-based surgical skin preps.

Authors:  Jason M Samuels; Heather Carmichael; Krzysztof J Wikiel; Thomas N Robinson; Carlton C Barnett; Teresa S Jones; Edward L Jones
Journal:  Surg Endosc       Date:  2019-06-20       Impact factor: 4.584

2.  Risks and prevention of surgical fires : A systematic review.

Authors:  I Kezze; N Zoremba; R Rossaint; A Rieg; M Coburn; G Schälte
Journal:  Anaesthesist       Date:  2018-05-15       Impact factor: 1.041

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.