Literature DB >> 24669600

Intraocular pressure measurement: Goldmann Applanation Tonometer vs non contact airpuff tonometer.

Masood Alam Shah, Khalid Bin Saleem, Talat Mehmood.   

Abstract

BACKGROUND: An accurate assessment of Intraocular pressure (IOP) is vital in establishing diagnosis of Glaucoma and decision making regarding various treatment modalities available. The purpose of this study is to compare Goldmann Applanation Tonometer (GAT) with Air puff tonometer.
METHODS: Cross-sectional comparative study conducted. 73 eyes from 73 patients were included in this study and intraocular pressure (IOP) was measured by GAT and PT100 at Sheikh Khalifa Bin Zayed Hospital, Muzaffarabad, Benazir Shaheed Teaching Hospital, Abbottabad.
RESULTS: Mean age of the patients was 53.17 +/- 13.80 years. Mean IOP measurements showed significant differences in measurements performed by the two tonometers (p < 0.05). Correlation revealed significant relation between PT100 and GAT (Pearson's correlation 0.715, p < 0.01).
CONCLUSIONS: Compared to non-contact air-puff tonometer, the Goldmann applanation tonometer is a reliable and consistent technique for measurement of intraocular pressure.

Entities:  

Mesh:

Year:  2012        PMID: 24669600

Source DB:  PubMed          Journal:  J Ayub Med Coll Abbottabad        ISSN: 1025-9589


  3 in total

1.  Effect of manual upper eyelid elevation on intraocular pressure measurement by noncontact tonometer.

Authors:  Raziye Dönmez Gün; Şaban Şimşek
Journal:  Int Ophthalmol       Date:  2022-05-13       Impact factor: 2.029

2.  Correlation Between Intraocular Pressure and Central Corneal Thickness in Persian Children.

Authors:  Mahmood Nejabat; Fatemeh Heidary; Mohammad Reza Talebnejad; Ramin Salouti; Mohammad Hossein Nowroozzadeh; Masoumeh Masoumpour; Hamideh Mahdaviazad; Zahra Tajbakhsh; Maryam Keshtkar; Hossein Jamali; Mohammad Reza Khalili; Hossein Movahedan; Narges Roustaei; Reza Gharebaghi
Journal:  Ophthalmol Ther       Date:  2016-10-05

3.  Comparison of Goldmann Applanation, Noncontact Air Puff, and Tono-Pen XL Tonometry in Normal Controls versus Glaucoma Patients at a University Hospital in Riyadh, Saudi Arabia.

Authors:  Essam A Osman; Priscilla W Gikandi; Abdul-Rahman A Al-Jasser; Majeedah Alotaibi; Ahmed Mousa
Journal:  Middle East Afr J Ophthalmol       Date:  2018 Jan-Mar
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.