| Literature DB >> 24668857 |
Wenyan Li1, Giacomo Vacca, Maryann Castillo, Kevin D Houston, Jessica P Houston.
Abstract
Flow cytometers are powerful high-throughput devices that capture spectroscopic information from individual particles or cells. These instruments provide a means of multi-parametric analyses for various cellular biomarkers or labeled organelles and cellular proteins. However, the spectral overlap of fluorophores limits the number of fluorophores that can be used simultaneously during experimentation. Time-resolved parameters enable the quantification of fluorescence decay kinetics, thus circumventing common issues associated with intensity-based measurements. This contribution introduces fluorescence lifetime excitation cytometry by kinetic dithering (FLECKD) as a method to capture multiple fluorescence lifetimes using a hybrid time-domain approach. The FLECKD approach excites fluorophores by delivering short pulses of light to cells or particles by rapid dithering and facilitates measurement of complex fluorescence decay kinetics by flow cytometry. Our simulations demonstrated a resolvable fluorescence lifetime value as low as 1.8 ns (±0.3 ns) with less than 20% absolute error. Using the FLECKD instrument, we measured the shortest average fluorescence lifetime value of 2.4 ns and found the system measurement error to be ±0.3 ns (SEM), from hundreds of monodisperse and chemically stable fluorescent microspheres. Additionally, we demonstrate the ability to detect two distinct excited state lifetimes from fluorophores in single cells using FLECKD. This approach presents a new ability to resolve multiple fluorescence lifetimes while retaining the fluidic throughput of a cytometry system. The ability to discriminate more than one average fluorescence lifetime expands the current capabilities of high-throughput and intensity-based cytometry assays as the need to tag one single cell with multiple fluorophores is now widespread.Entities:
Keywords: Bioanalytical; Flow cytometry; Fluorescence decay; Fluorescence lifetime; Time-resolved
Mesh:
Substances:
Year: 2014 PMID: 24668857 PMCID: PMC4231566 DOI: 10.1002/elps.201300618
Source DB: PubMed Journal: Electrophoresis ISSN: 0173-0835 Impact factor: 3.535
Figure 1Illustration of a laser beam waist with a Gaussian intensity profile interacting with a cell (circular shape). As a cell enters the beam at 1, is fully illuminated at 2, and leaves the beam at 3, fluorescence emission (and corresponding light scatter) increases from 0 to its peak value and then decreases to 0 (bottom). The finite fluorescence lifetime causes a time delay Δτ of the fluorescence signal.
Figure 2Schematic of the fluorescence lifetime excitation cytometry by kinetic dithering (FLECKD) system.
Figure 3Sets of simulations performed to model side scatter and fluorescence waveforms. (A), (B), (C) Gaussian functions (—) representing side-scatter waveforms with 1.5 μs, 15 ns, and 15 ps FWHM, respectively. The ex-Gaussian curves (o) were obtained by convolving each Gaussian function with a 10-ns exponential decay. (D) Ex-Gaussian curves produced by a Gaussian function (15-ns FWHM) convolved with 1− (—), 6-(o), 11− (*), 16− (x), 21− (□), 26− (—) and 31-ns (▽) exponential decays. (E) Ex-Gaussian curves produced by a Gaussian function (15-ns FWHM) convolved with multi-exponential decays (A1 = 0.7, τ1 = 2 ns; A2 = 0.3, τ2 = 15 ns) (—), (A1 = 0.3, τ1 = 2 ns; A2 = 0.7, τ2 = 15 ns) (—), (A1 = 0.7, τ1 = 8 ns; A2 = 0.3, τ2 = 15 ns) (*), and (A1 = 0.7, τ1 = 2 ns; A2 = 0.3, τ2 = 22 ns) (x), respectively. (F) Two ex-Gaussian curves obtained by a convolution of a Gaussian function (15-ns FWHM) with a double-exponential decay function (A1 = 0.7, τ1 = 2 ns; A2 = 0.3, τ2 = 22 ns) (—) and mean lifetime of 18.5 ns (70% 2 ns and 30% 22 ns) (—).
Summary of the simulation results for a variety of fluorescence waveforms with different fluorescence lifetimes
| True lifetime (ns) | Calculated lifetime (ns) | Error (%) | True lifetime (ns) | Calculated lifetime (ns) | Error (%) |
|---|---|---|---|---|---|
| 0.5 | 0.17 | 66 | 2.1 | 1.8 | 13 |
| 0.7 | 0.34 | 51 | 2.2 | 1.8 | 17 |
| 1 | 0.62 | 37 | 2.5 | 2.3 | 8 |
| 1.3 | 0.93 | 29 | 3 | 2.8 | 6 |
| 1.4 | 0.93 | 34 | 4 | 3.7 | 8 |
| 1.5 | 1 | 32 | 5 | 4.7 | 7 |
| 1.6 | 1.2 | 25 | 6 | 5.7 | 5 |
| 1.7 | 1.4 | 18 | 7 | 6.8 | 2 |
| 1.8 | 1.4 | 22 | 8 | 7.6 | 5 |
| 1.9 | 1.6 | 16 | 10 | 9.7 | 3 |
| 2 | 1.6 | 20 |
The ‘true lifetime’ column provides simulated data using Eq. (3) by a convolution of a Gaussian function with random noise and single-exponential decay functions. The fit of each simulated fluorescence waveform was determined by finding the optimum lifetime (noted as ‘calculated lifetime’) at minimum fitting error. The calculated lifetimes are listed in the second column. The absolute error percentages between the true lifetimes and the calculated lifetimes are listed in the third column.
Figure 4Experimental data waveforms (side scatter, —) (Fluorescence, —*—) and fitting curves (Fitting, —o—) were obtained by measuring fluorescence particles and EB-stained CHO K1 cells with the FLECKD system. (A) Nile Red particles, (B) Flow-Check Fluorospheres™, and (C) DNA-bound EB. Panel (D) is a distribution of fluorescence lifetime values obtained from cells expressing tdTomato.
This table summarizes the fluorescence lifetime results from cell measurements in the presence and absence of amsacrine, the fluorophore quencher
| 0 μM | 18.6 μM | 27.9 μM | 37.2 μM | |
|---|---|---|---|---|
| EB intensity percentage decrease | 0% | 2% | 19% | 24% |
| EB mean lifetime (tau1) | 15.9 ns | 15.2 ns | 14.6 ns | 14.0 ns |
| EB mean lifetime (tau2) | 0.9 ns | 0.9 ns | 0.7 ns | 1.3 ns |
| Mean reduced-Chi-square errors | 3.88 | 4.34 | 2.9 | 4.1 |
The two-lifetime components determined are listed as the reduced mean χ2 for the comparison between experiment and model. The average decrease in fluorescence intensity is also provided.