Literature DB >> 24666841

Myocardial hydrophilic polymer emboli following cardiac catheterization: a case report and literature review.

Lauren E Rosen1, Rohit I Singh2, Brett Mahon2.   

Abstract

BACKGROUND: Intravascular polymer emboli have been reported in the skin, lungs, and brain following vascular procedures utilizing hydrophilic polymer coated devices. The Cook arterial introducer sheath was the first of these devices to be introduced, after which case reports followed documenting sterile inflammation at the sheath access site, characterized histologically by perivascular granulomas containing hydrophilic polymer. More recently, hydrophilic polymer emboli have been reported in the vessels of the lungs and brain in association with ischemia and infarct following vascular procedures using polymer coated devices.
METHODS: We report a case of intravascular myocardial emboli associated with acute myocardial infarction following cardiac catheterization. The patient was a 65-year-old man who received cardiac catheterization with placement of a bare metal stent following myocardial infarction. One month later, the patient presented with angina and died shortly after admission.
RESULTS: Autopsy revealed hemopericardium with rupture of the left ventricle through an aneurismal defect within the area of prior infarction. Microscopically, an area of acute infarction was present within the aneurismal defect. Numerous small and medium-sized vessels within the left ventricle were occluded by basophilic amorphous granular material with an inflammatory giant cell response.
CONCLUSION: The emboli were most frequent in the area of acute infarction, suggesting that the emboli may have resulted in ischemia leading to the patient's death. This is the third documented case of intramyocardial polymer emboli following cardiac catheterization and the first case to our knowledge to document an association between intravascular myocardial polymer emboli and acute myocardial infarction.
Copyright © 2014 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Cardiac catheterization; Hydrophilic polymer emboli; Myocardial infarction

Mesh:

Substances:

Year:  2014        PMID: 24666841     DOI: 10.1016/j.carpath.2014.01.009

Source DB:  PubMed          Journal:  Cardiovasc Pathol        ISSN: 1054-8807            Impact factor:   2.185


  4 in total

Review 1.  Understanding the Basics of Commonly Used Wires in Interventional Radiology.

Authors:  Mithil B Pandhi; Ravi Tyagi; Divya Surabhi; Ketan Y Shah
Journal:  Semin Intervent Radiol       Date:  2021-10-07       Impact factor: 1.780

Review 2.  Non-ischemic cerebral enhancing lesions secondary to endovascular aneurysm therapy: nickel allergy or foreign body reaction? Case series and review of the literature.

Authors:  Eimad Shotar; Bruno Law-Ye; Flore Baronnet-Chauvet; Sinead Zeidan; Dimitri Psimaras; Franck Bielle; Catherine Pecquet; Soledad Navarro; Charlotte Rosso; Fleur Cohen; Jacques Chiras; Federico Di Maria; Nader Sourour; Frédéric Clarençon
Journal:  Neuroradiology       Date:  2016-05-23       Impact factor: 2.804

3.  Cerebral Granulomatous Inflammation Secondary to Hydrophilic Polymer Embolization Following Thrombectomy.

Authors:  Craig Meiers; Yoftahe Abebe; Neville M Alberto; John Riedinger; Dane A Breker; Michael Manchak; Alexander Drofa; Corey Teigen
Journal:  Am J Case Rep       Date:  2017-05-08

Review 4.  Hydrophilic Polymer Embolism: Implications for Manufacturing, Regulation, and Postmarket Surveillance of Coated Intravascular Medical Devices.

Authors:  Rashi I Mehta; Rupal I Mehta
Journal:  J Patient Saf       Date:  2021-12-01       Impact factor: 2.844