Asgeir S Jakola1, Arve Jørgensen, Tormod Selbekk, Ralf-Peter Michler, Ole Solheim, Sverre H Torp, Lisa M Sagberg, Petter Aadahl, Geirmund Unsgård.
Abstract
BACKGROUND: Use of ultrasound in brain tumor surgery is common. The difference in attenuation between brain and isotonic saline may cause artifacts that degrade the ultrasound images, potentially affecting resection grades and safety. Our research group has developed an acoustic coupling fluid that attenuates ultrasound energy like the normal brain. We aimed to test in animals if the newly developed acoustic coupling fluid may have harmful effects.Entities:
Mesh:
Substances:
Year: 2014 PMID: 24666721 PMCID: PMC3977940 DOI: 10.1186/1471-2342-14-11
Source DB: PubMed Journal: BMC Med Imaging ISSN: 1471-2342 Impact factor: 1.930
Figure 1Ultrasound imaging during glioma resection. Screen dump from the ultrasound based neuronavigation system (SonoWand®). Images to the left are preoperative MRI demonstrating a hyperintense region suspect of diffuse glioma (above; sagittal view, below; coronal view). Images in the middle represent the corresponding ultrasound images taken intraoperatively before resection. Pictures to the right demonstrate corresponding ultrasound images approximately midway in the resection. Notice the bright-rim in the bottom of the cavity, representing an artifact, making the interpretation of hyperintense areas that could represent tumor on the images somewhat more difficult.
Figure 2EEG registrations immediate after the first injection of penicillin G leading to generalized rhythmic fast sharp activity.
Figure 3EEG registrations after the second injection of penicillin G leading to seizure activity.
Scores of symptoms in pigs post-procedure
| 0/12 | 0/12 | 1/12 | 1/12 | 1/12 | 1/12 | |
| 0/12 | 1/12 | 0/12 | 0/12 | 0/12 | 0/12 | |
| 0/12 | 0/12 | 0/12 | 0/12 | 0/12 | 0/12 | |
| 1/12 | 2/12 | 0/12 | 0/12 | 0/12 | 0/12 | |
| 0/12 | 0/12 | 0/12 | 0/12 | 0/12 | 0/12 | |
| 2/12 | 2/12 | 1/12 | 0/12 | 0/12 | 0/12 |
The score was predefined by a veterinarian and the score was based on: appearance, behavior, nutritional status and relevant clinical signs (gait disturbances and/or seizure). For each dimension a range from 0 (normal) to 3 (markedly abnormal) was possible giving a range from 0-12 on each observation. The scores presented were from the assessment where the individual pig reached the highest sum (most symptoms). A total score of 0-3 was predefined as normal unless the pig scored 3 in one dimension (severe symptoms). A total score higher than 3 or maximal score in one dimension warranted a consult with a veterinarian.
Figure 4EEG registrations in pig 3. a) Before the procedure. Synchronous background activity 4-6 Hz with highest amplitudes in the frontotemporal region. b) After injection of the acoustic coupling fluid. There was no evidence of seizure activity on EEG.
Figure 5Histology of injection site of the acoustic fluid in rat. There was a minor macrophage response just around the needle site which is a normal reaction to brain trauma. Except from this there was no evidence of any reactions around the needle site.