MR arthrography: impact of steroids, local anesthetics, and iodinated contrast material on gadolinium signal intensity in phantoms at 1.5 and 3.0 T.

Marco A Ugas1, Bang H Huynh, Michael G Fox, James T Patrie, Cree M Gaskin.   

Abstract

PURPOSE: To determine if gadolinium signal intensity at direct magnetic resonance (MR) arthrography is affected by the addition of steroids, anesthetics, or iodinated contrast material.
MATERIALS AND METHODS: This study did not require approval by the institutional review board because no patients or patient data were involved. An in vitro study was performed to evaluate various concentrations of three gadolinium contrast agents (gadopentetate dimeglumine, gadobenate dimeglumine, and gadofosveset trisodium) diluted in either saline or iodinated contrast material (50% and full-strength iohexol 300). Three steroids (betamethasone, triamcinolone, and methylprednisolone) and three local anesthetics (lidocaine, ropivacaine, and bupivacaine) were added to solutions in clinical doses. T1-weighted fat-suppressed MR imaging sequences were performed in phantoms at 1.5 and 3.0 T. Signal intensities were measured. All experiments were repeated in full for a total of three replications each. The data were analyzed by using two-way factorial analysis of variance.
RESULTS: Dilution of gadolinium into iohexol reduced the signal intensity in all samples compared with dilution in saline alone. Peak signal intensities were at 0.625 and 1.25 mmol/L of gadolinium in iohexol at both magnet strengths. At 1.5 T, the addition of steroids and anesthetics to the saline solutions had no impact on the signal intensity curves, with the peak signal intensity at gadolinium concentrations of 2.5 and 1.25 mmol/L. At 3.0 T, the addition of steroids and anesthetics had minimal effect on signal intensity curves, with the peak signal intensity at 1.25 mmol/L of gadolinium.
CONCLUSION: The addition of steroids and/or anesthetics to gadolinium solutions for MR arthrography does not substantially impact signal intensity. When gadolinium is diluted into a 50% or greater strength of iohexol, the signal intensity curve shifts so that the maximum signal intensity is obtained with lower gadolinium concentrations (0.625-1.25 mmol/L).

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24661248     DOI: 10.1148/radiol.14132352

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  4 in total

1.  Ultrasound guidance to perform intra-articular injection of gadolinium-based contrast material for magnetic resonance arthrography as an alternative to fluoroscopy: the time is now.

Authors:  Carmelo Messina; Giuseppe Banfi; Alberto Aliprandi; Giovanni Mauri; Francesco Secchi; Francesco Sardanelli; Luca Maria Sconfienza
Journal:  Eur Radiol       Date:  2015-08-08       Impact factor: 5.315

2.  Diagnostic performance of shoulder magnetic resonance arthrography for labral tears having surgery as reference: comparison of high-resolution isotropic 3D sequence (THRIVE) with standard protocol.

Authors:  Sun Hwa Lee; Seong Jong Yun; Youngno Yoon
Journal:  Radiol Med       Date:  2018-03-26       Impact factor: 3.469

Review 3.  Gadolinium deposition and the potential for toxicological sequelae - A literature review of issues surrounding gadolinium-based contrast agents.

Authors:  Kerry A Layne; Paul I Dargan; John R H Archer; David M Wood
Journal:  Br J Clin Pharmacol       Date:  2018-08-17       Impact factor: 4.335

4.  Effects of adding local anesthetic and iodinated contrast agents to the paramagnetic contrast solution in direct MR arthrography.

Authors:  Marcello Henrique Nogueira-Barbosa
Journal:  Radiol Bras       Date:  2015 Mar-Apr
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.