Effect of mattress and bed frame deflection on real chest compression depth measured with two CPR sensors.

Marko Sainio1, Heidi Hellevuo2, Heini Huhtala3, Sanna Hoppu4, Joar Eilevstjønn5, Jyrki Tenhunen6, Klaus T Olkkola7.   

Abstract

AIM: Implementation of chest compression (CC) feedback devices with a single force and deflection sensor (FDS) may improve the quality of CPR. However, CC depth may be overestimated if the patient is on a compliant surface. We have measured the true CC depth during in-hospital CPR using two FDSs on different bed and mattress types.
METHODS: This prospective observational study was conducted at Tampere University Hospital between August 2011 and September 2012. During in-hospital CPR one FDS was placed between the rescuer's hand and the patient's chest, with the second attached to the backboard between the patient's back and the mattress. The real CC depth was calculated as the difference between the total depth from upper FDS to lower FDS.
RESULTS: Ten cardiac arrests on three different bed and mattress types yielded 10,868 CCs for data analyses. The mean (SD) mattress/bed frame effect was 12.8 (4) mm on a standard hospital bed with a gel mattress, 12.4 (4) mm on an emergency room stretcher with a thin gel mattress and 14.1 (3) mm on an ICU bed with an emptied air mattress. The proportion of CCs with an adequate depth (≥50 mm) decreased on all mattress types after compensating for the mattress/bed frame effect from 94 to 64%, 98 to 76% and 91 to 17%, in standard hospital bed, emergency room stretcher and ICU bed, respectively (p<0.001).
CONCLUSION: The use of FDS without real-time correction for deflection may result in CC depth not reaching the recommended depth of 50 mm.
Copyright © 2014 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Keywords:  Beds; Cardiopulmonary resuscitation; Feedback; Mattresses; Resuscitation; Sensor

Mesh:

Year:  2014        PMID: 24657249     DOI: 10.1016/j.resuscitation.2014.03.009

Source DB:  PubMed          Journal:  Resuscitation        ISSN: 0300-9572            Impact factor:   5.262


  3 in total

1.  Performance of different support surfaces during experimental resuscitation (CPR).

Authors:  Esa Soppi; Ansa Iivanainen; Leila Sikanen; Elina Jouppila-Kupiainen
Journal:  Heliyon       Date:  2016-02-19

2.  Reducing the impact of intensive care unit mattress compressibility during CPR: a simulation-based study.

Authors:  Yiqun Lin; Brandi Wan; Claudia Belanger; Kent Hecker; Elaine Gilfoyle; Jennifer Davidson; Adam Cheng
Journal:  Adv Simul (Lond)       Date:  2017-11-16

3.  Do automated real-time feedback devices improve CPR quality? A systematic review of literature.

Authors:  Debora Gugelmin-Almeida; Lucia Tobase; Thatiane Facholi Polastri; Heloisa Helena Ciqueto Peres; Sergio Timerman
Journal:  Resusc Plus       Date:  2021-03-27
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.