Literature DB >> 24656790

Use of CT dose notification and alert values in routine clinical practice.

Michelle E Howard1, Cynthia H McCollough2, Shuai Leng1, Lifeng Yu1, Mike R Bruesewitz1.   

Abstract

OBJECTIVE: The purpose of this investigation was to assess the impact on workflow of the use of notification and alert values in our practice and to provide baseline data for quality improvement initiatives. METHODS AND MATERIALS: Five diagnostic clinical CT scanners were programmed with the notification and alert values recommended by the American Association of Physics in Medicine. Retrospective analysis was performed on log files to assess the frequency of and reason for notification and alert events.
RESULTS: Between February and September of 2012, 11,384 patients were scanned on the 5 systems. One alert occurred because of the use of bolus tracking in a morbidly obese patient, where the prescan cumulative volume CT dose index for the exam exceeded the recommended alert value of 1,000 mGy. Only 1.2 ± 0.6% of patient scans triggered a notification. Notifications were mainly triggered because of bolus tracking and/or large patient size. Protocols triggering notifications most often included CT angiography of the chest for pulmonary emboli.
CONCLUSION: Because only a small percentage of performed patient examinations triggered a notification or alert event, the impact on workflow of adopting these features was negligible. Evaluation of the logs identified trends in reasons for which notification events were triggered; these primarily included large patient size and bolus tracking. Additionally, specific protocols were identified where adjustment of notification values was deemed necessary.
Copyright © 2014 American College of Radiology. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  CT; dose check; notification and alert values; patient safety initiatives; radiation dose

Mesh:

Year:  2014        PMID: 24656790     DOI: 10.1016/j.jacr.2013.12.017

Source DB:  PubMed          Journal:  J Am Coll Radiol        ISSN: 1546-1440            Impact factor:   5.532


  3 in total

Review 1.  Notifications and alerts in patient dose values for computed tomography and fluoroscopy-guided interventional procedures.

Authors:  Eliseo Vano; Reinhard Loose; Guy Frija; Graciano Paulo; Efstathios Efstathopoulos; Claudio Granata; Riccardo Corridori; Alberto Torresin; Jonas S Andersson; Virginia Tsapaki; Josefin Ammon; Christoph Hoeschen
Journal:  Eur Radiol       Date:  2022-03-16       Impact factor: 7.034

2.  Three-Dimensional Reconstruction Modeling of the Spatial Displacement, Extent and Rotational Orientation of Undisplaced Femoral Neck Fractures.

Authors:  Xin Fu; Gui-Jun Xu; Zhi-Jun Li; Chang-Ling Du; Zhe Han; Tao Zhang; Xinlong Ma
Journal:  Medicine (Baltimore)       Date:  2015-09       Impact factor: 1.817

3.  Standard Anatomic Terminologies: Comparison for Use in a Health Information Exchange-Based Prior Computed Tomography (CT) Alerting System.

Authors:  Anton Oscar Beitia; Tina Lowry; Daniel J Vreeman; George T Loo; Bradley N Delman; Frederick L Thum; Benjamin H Slovis; Jason S Shapiro
Journal:  JMIR Med Inform       Date:  2017-12-14
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.