The effect of metallic tracheal stents on radiation dose in the airway and surrounding tissues.

Andrew J Evans1, David Y Lee2, Anudh K Jain1, Syed S Razi2, Koji Park2, Gary S Schwartz2, Frieda Trichter1, Jason Ostenson1, Jordan R Sasson3, Faiz Y Bhora2.   

Abstract

BACKGROUND: Metallic airway stents are often used in the management of central airway malignancies. The presence of a metallic foreign body may affect radiation dose in tissue. We studied the effect of a metallic airway stent on radiation dose delivery in a phantom and an in vivo porcine model.
METHODS: A metallic tracheal stent was fitted onto a support in a water phantom. Point dosimeters were positioned in the phantom around the support and the stent. Irradiation was then performed on a linear accelerator with and without the stent. Metallic tracheal stents were deployed in the trachea of three pigs. Dosimeters were implanted in the tissues near (Group 1) and away (Group 2) from the stent. The pigs were then irradiated, and the dose perturbation factor was calculated by comparing the actual dose detected by the dosimeters versus the planned dose.
RESULTS: The difference in the dose detected by the dosimeters and the planned dose ranged from 1.8% to 6.1% for the phantom with the stent and 0%-5.3% for the phantom without the stent. These values were largely within the manufacturer's specified error of 5%. No significant difference was observed in the dose perturbation factor for Group 1 and Group 2 dosimeters (0.836 ± 0.058 versus 0.877 ± 0.088, P = 0.220) in all the three pigs.
CONCLUSIONS: Metallic airway stents do not significantly affect radiation dose in the airway and surrounding tissues in a phantom and porcine model. Radiation treatment planning systems can account for the presence of the stent. External beam radiation can be delivered without concern for significant dose perturbation.
Copyright © 2014. Published by Elsevier Inc.

Entities:  

Keywords:  Airway malignancy; Airway obstruction; Bronchial malignancy; Radiation; Stents

Mesh:

Substances:

Year:  2014        PMID: 24656475     DOI: 10.1016/j.jss.2014.01.013

Source DB:  PubMed          Journal:  J Surg Res        ISSN: 0022-4804            Impact factor:   2.192


  2 in total

1.  Assessment of dose perturbations for metal stent in photon and proton radiotherapy plans for hepatocellular carcinoma.

Authors:  Boram Lee; Sungkoo Cho; Hee Chul Park; Sang-Won Kang; Jae-Sung Kim; Jin-Beom Chung
Journal:  Radiat Oncol       Date:  2022-07-16       Impact factor: 4.309

2.  A new removable airway stent.

Authors:  Tore Amundsen; Sveinung Sørhaug; Håkon Olav Leira; Stig Sverre Tyvold; Thomas Langø; Tommy Hammer; Frode Manstad-Hulaas; Erney Mattsson
Journal:  Eur Clin Respir J       Date:  2016-09-06