Evaluation of the sealing capability of implants to titanium and zirconia abutments against Porphyromonas gingivalis, Prevotella intermedia, and Fusobacterium nucleatum under different screw torque values.

Nicole A Smith1, Ilser Turkyilmaz2.   

Abstract

STATEMENT OF PROBLEM: When evaluating long-term implant success, clinicians have always been concerned with the gap at the implant-abutment junction, where bacteria can accumulate and cause marginal bone loss. However, little information regarding bacterial leakage at the implant-abutment junction, or microgap, is available.
PURPOSE: The purpose of this study was to evaluate sealing at 2 different implant-abutment interfaces under different screw torque values.
MATERIAL AND METHODS: Twenty sterile zirconia abutments and 20 sterile titanium abutments were screwed into 40 sterile implants and placed in test tubes. The ability of a bacterial mixture of Prevotella intermedia, Porphyromonas gingivalis, and Fusobacterium nucleatum to leak through an implant-titanium abutment seal under 20 and 35 Ncm torque values and an implant-zirconia abutment seal under 20 and 35 Ncm torque values was evaluated daily until leakage was noted. Once a unit demonstrated leakage, a specimen was plated. After 4 days, the number of colonies on each plate was counted with an electronic colony counter. Plating was used to verify whether or not bacterial leakage occurred and when leakage first occurred. The implant-abutment units were removed and rinsed with phosphate buffered saline solution and evaluated with a stereomicroscope. The marginal gap between the implant and the abutment was measured and correlated with the amount of bacterial leakage. The data were analyzed with ANOVA.
RESULTS: Bacterial leakage was noted in all specimens, regardless of material or screw torque value. With titanium abutments, changing the screw torque value from 20 to 35 Ncm did not significantly affect the amount of bacterial leakage. However, with zirconia abutments, changing the screw torque value from 20 to 35 Ncm was statistically significant (P<.017). Overall, the marginal gap noted was larger at the zirconia-abutment interface (5.25 ±1.99 μm) than the titanium-abutment interface (12.38 ±3.73 μm), irrespective of the screw torque value. Stereomicroscopy revealed a nonuniform marginal gap in all specimens.
CONCLUSION: The results of this study showed that, over time, bacteria will leak through the implant-abutment microgap at the implant-abutment interface. Implants with a titanium abutment demonstrate a smaller microgap than implants with a zirconia abutment. Tightening the zirconia abutment screw from 20 to 35 Ncm decreases the size of the microgap, which suggests a more intimate fit between the implant and the abutment.
Copyright © 2014 Editorial Council for the Journal of Prosthetic Dentistry. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24656409     DOI: 10.1016/j.prosdent.2013.11.010

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  10 in total

Review 1.  Microleakage at the Different Implant Abutment Interface: A Systematic Review.

Authors:  Sunil Kumar Mishra; Ramesh Chowdhary; Shail Kumari
Journal:  J Clin Diagn Res       Date:  2017-06-01

2.  Microgap and microleakage of a hybrid connection platform-switched implant system in the absence or presence of a silicone-based sealing agent.

Authors:  Pedro Ferrás Fernandes; Liliana Grenho; Maria Helena Fernandes; João Carlos Sampaio-Fernandes; Pedro Sousa Gomes
Journal:  Odontology       Date:  2021-08-27       Impact factor: 2.634

3.  Effect of Fabrication Technique on the Microgap of CAD/CAM Cobalt-Chrome and Zirconia Abutments on a Conical Connection Implant: An In Vitro Study.

Authors:  Pedro Molinero-Mourelle; Rocio Cascos-Sanchez; Burak Yilmaz; Walter Yu Hang Lam; Edmond Ho Nang Pow; Jaime Del Río Highsmith; Miguel Gómez-Polo
Journal:  Materials (Basel)       Date:  2021-04-30       Impact factor: 3.623

Review 4.  Microbial Profiles and Detection Techniques in Peri-Implant Diseases: a Systematic Review.

Authors:  Miguel Padial-Molina; Jesús López-Martínez; Francisco O'Valle; Pablo Galindo-Moreno
Journal:  J Oral Maxillofac Res       Date:  2016-09-09

5.  Effects of Liner-Bonding of Implant-Supported Glass-Ceramic Crown to Zirconia Abutment on Bond Strength and Fracture Resistance.

Authors:  Yong-Seok Jang; Sang-Hoon Oh; Won-Suck Oh; Min-Ho Lee; Jung-Jin Lee; Tae-Sung Bae
Journal:  Materials (Basel)       Date:  2019-08-30       Impact factor: 3.623

6.  Sealing agent reduces formation of single and dual-species biofilms of Candida albicans and Enterococcus faecalis on screw joints at the abutment/implant interface.

Authors:  Cecília Alves de Sousa; Jadison Junio Conforte; Karina Sampaio Caiaffa; Cristiane Duque; Wirley Gonçalves Assunção
Journal:  PLoS One       Date:  2019-10-22       Impact factor: 3.240

7.  Influence of Implant Connection, Abutment Design and Screw Insertion Torque on Implant-Abutment Misfit.

Authors:  Jorge Vélez; Jesús Peláez; Carlos López-Suárez; Rubén Agustín-Panadero; Celia Tobar; María J Suárez
Journal:  J Clin Med       Date:  2020-07-24       Impact factor: 4.241

8.  Bacterial translocation and microgap formation at a novel conical indexed implant abutment system for single crowns.

Authors:  Peter Gehrke; Simon Burg; Ulrike Peters; Thomas Beikler; Carsten Fischer; Frank Rupp; Ernst Schweizer; Paul Weigl; Robert Sader; Ralf Smeets; Sogand Schäfer
Journal:  Clin Oral Investig       Date:  2021-08-16       Impact factor: 3.573

9.  The Use of the Diode Laser against the Microbiome on Composites Closing the Screw Access Hall (Sah) in the Reconstruction of Dental Implants: Ex Vivo Studies.

Authors:  Anna Wawrzyk; Mansur Rahnama; Weronika Sofińska-Chmiel; Sławomir Wilczyński; Michał Łobacz
Journal:  Int J Environ Res Public Health       Date:  2022-06-18       Impact factor: 4.614

10.  Evaluation of sealing efficacy and removal convenience of sealing materials for implant abutment screw access holes.

Authors:  Huangjun Zhou; Sixian Ye; Xingyu Lyu; Hao Feng; Min Liu; Cai Wen
Journal:  BMC Oral Health       Date:  2022-08-25       Impact factor: 3.747

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.