Literature DB >> 24655001

Comparing optic nerve head analysis between confocal scanning laser ophthalmoscopy and spectral domain optical coherence tomography.

Gloria Roberti1, Marco Centofanti, Francesco Oddone, Lucia Tanga, Manuele Michelessi, Gianluca Manni.   

Abstract

PURPOSE/AIM OF THE STUDY: Confocal scanning laser ophthalmoscopy, HRT3, and spectral domain optical coherence tomography (OCT), RTVue-100, are able to give 3-dimensional (3D) topography images of optic nerve head (ONH) and to derive stereometric parameters and sectorial analysis. The purpose of the study is to evaluate the agreement of these two devices and their diagnostic accuracy to discriminate eyes with glaucoma from those without.
MATERIALS AND METHODS: Glaucoma patients and healthy control subjects were included. All of them underwent a complete ophthalmological examination, including slit lamp evaluation and visual field (VF) test. After pupil dilatation, HRT3 and RTVue-100 were performed. The following stereometric parameters were recorded: disc area, rim area, rim volume, cup volume, cup area, cup/disk ratio, and the following sectors, superotemporal, superonasal, inferotemporal, inferonasal.
RESULTS: Forty-six eyes of 46 glaucoma patients and 58 eyes of 58 healthy subjects were included in the study. In both groups, HRT3 rim area and rim volume were statistically higher than RTVue-100 (glaucomas: 0.95 ± 0.38 versus 0.44 ± 0.33 and 0.19 ± 0.13 versus 0.02 ± 0.03, p < 0.01. controls: 1.41 ± 0.30 versus 1.08 ± 0.37 and 0.37 ± 0.13 versus 0.14 ± 0.11, p < 0.01), while cup area was statistically higher by RTVue-100 (glaucomas: 1.42 ± 0.57 versus 1.14 ± 0.58, p < 0.01. controls: 1.05 ± 1.35 versus 0.65 ± 0.48). Bland and Altman plots confirmed the presence of a fixed bias. The parameters with largest AUROC were rim volume, rim area and cup/disk ratio for both instruments. HRT3 inferotemporal sector had the highest sensitivity (80.43%, at 75.9% specificity), while for RTVue-100, the superotemporal sector had the highest sensitivity (76.1%, at 81% specificity). The agreement was moderate for inferotemporal sector and fair for the others.
CONCLUSIONS: HRT3 and RTVue-100 are not interchangeable for ONH analysis. They both have good diagnostic accuracy, but RTVue-100 shows slightly better performance, at least with regard to rim area volume.

Entities:  

Keywords:  Glaucoma; HRT3; Rtvue-100; imaging; optic nerve head

Mesh:

Year:  2014        PMID: 24655001     DOI: 10.3109/02713683.2014.891752

Source DB:  PubMed          Journal:  Curr Eye Res        ISSN: 0271-3683            Impact factor:   2.424


  3 in total

Review 1.  Optic nerve head and fibre layer imaging for diagnosing glaucoma.

Authors:  Manuele Michelessi; Ersilia Lucenteforte; Francesco Oddone; Miriam Brazzelli; Mariacristina Parravano; Sara Franchi; Sueko M Ng; Gianni Virgili
Journal:  Cochrane Database Syst Rev       Date:  2015-11-30

2.  Comparison of Diagnostic Ability Between Wide-Field Swept-Source Optical Coherence Tomography Imaging Maps and Heidelberg Retina Tomograph 3 Optic Nerve Head Assessment to Discriminate Glaucomatous and Non-glaucomatous Eyes.

Authors:  Dimitrios Kourkoutas; George Triantafyllopoulos; Iordanis Georgiou; Aristotelis Karamaounas; Nikolaos Karamaounas; Konstadinos Sotiropulos; Dimitrios Kapralos
Journal:  Cureus       Date:  2022-08-19

3.  Optic disc morphology in unilateral branch retinal vein occlusion using spectral domain optical coherence tomography.

Authors:  Andrea Szigeti; Miklós Schneider; Mónika Ecsedy; Zoltán Zs Nagy; Zsuzsanna Récsán
Journal:  BMC Ophthalmol       Date:  2015-12-12       Impact factor: 2.209

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.