Experiences with the use of iteratively reconstructed dose-modified MDCT angiography examinations of living renal donors.

Yasir Andrabi1, Avinash Kambadakone, Dushyant V Sahani.   

Abstract

PURPOSE: To evaluate the performance of iteratively reconstructed (IR) dose-modified (DM) multidetector computed tomography (MDCT) angiography (CTA) examinations of renal donors in comparison to standard dose filtered back projection (FBP) images.
MATERIALS AND METHODS: Eighty-five potential donors who underwent dual-phase CTA on 16-/64-MDCT scanners were retrospectively reviewed. Images from a 16-MDCT scanner were reconstructed with filtered back projection (group A, 47) and examinations from 64-MDCT with IR (group B, 38). Scan parameters were constant for both groups except for higher noise index (× 1.3) in group B. Images were interpreted for the relevant anatomy and IQ by 2 readers. Surgical report served as reference standard for operated kidneys, whereas for nonoperated kidneys, interobserver agreement was evaluated.
RESULTS: Radiation dose was 36% lower in group B compared to group A. All CTA examinations were rated for diagnostic quality with comparable IQ scores. In 48 operated kidneys, 10 surgically confirmed vascular anomalies were correctly identified by both readers. In the remaining 122 nonoperated kidneys, there was an excellent interobserver agreement.
CONCLUSIONS: Iteratively reconstructed technique preserves high image quality and diagnostic performance at significantly lower radiation doses in DM kidney donor CTA examinations.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24651752     DOI: 10.1097/RCT.0000000000000084

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  2 in total

1.  An augmented patient-specific approach to administration of contrast agent for CT renal angiography.

Authors:  Charbel Saade; Nadine Hamieh; Ibrahim Al-Sheikh Deeb; Maurice Haddad; Alain S Abi-Ghanem; Diamond Ghieh; Fadi El-Merhi
Journal:  Int Braz J Urol       Date:  2019 Sep-Oct       Impact factor: 1.541

2.  Renal Collecting System Anatomy in Living Kidney Donors by Computed Tomographic Urography: Protocol Accuracy Compared to Intravenous Pyelographic and Surgical Findings.

Authors:  Mohammad Kazem Tarzamni; Nariman Nezami; Afshar Zomorrodi; Samad Fathi-Noroozlou; Reza Piri; Mohammad Naghavi-Behzad; Mohammad Khalid Mojadidi; Bijan Bijan
Journal:  J Clin Imaging Sci       Date:  2016-01-28
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.