Literature DB >> 24650798

Evaluation and performance characteristics of the Q Hemostasis Analyzer, an automated coagulation analyzer.

Pierre Toulon1, Florence Fischer2, Anny Appert-Flory2, Didier Jambou2.   

Abstract

The Q Hemostasis Analyzer (Grifols, Barcelona, Spain) is a fully-automated random-access multiparameter analyzer, designed to perform coagulation, chromogenic and immunologic assays. It is equipped with a cap-piercing system. The instrument was evaluated in a hemostasis laboratory of a University Hospital with respect to its technical features in the determination of coagulation i.e. prothrombin time (PT), activated partial thromboplastin time (aPTT), thrombin time, fibrinogen and single coagulation factors V (FV) and VIII (FVIII), chromogenic [antithrombin (AT) and protein C activity] and immunologic assays [von Willebrand factor antigen (vWF:Ag) concentration], using reagents from the analyzer manufacturer. Total precision (evaluated as the coefficient of variation) was below 6% for most parameters both in normal and in pathological ranges, except for FV, FVIII, AT and vWF:Ag both in the normal and pathological samples. No carryover was detected in alternating aPTT measurement in a pool of normal plasma samples and in the same pool spiked with unfractionated heparin (>1.5 IU/mL). The effective throughput was 154 PT, 66 PT/aPTT, 42 PT/aPTT/fibrinogen, and 38 PT/aPTT/AT per hour, leading to 154 to 114 tests performed per hour, depending of the tested panel. Test results obtained on the Q Hemostasis Analyzer were well correlated with those obtained on the ACL TOP analyzer (Instrumentation Laboratory), with r between 0.862 and 0.989. In conclusion, routine coagulation testing can be performed on the Q Hemostasis Analyzer with satisfactory precision and the same apply to more specialized and specific tests.
Copyright © 2014 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Coagulation analyzer; Evaluation; Q Hemostasis Analyzer

Mesh:

Year:  2014        PMID: 24650798     DOI: 10.1016/j.thromres.2014.02.021

Source DB:  PubMed          Journal:  Thromb Res        ISSN: 0049-3848            Impact factor:   3.944


  1 in total

1.  From Hemophilia to Deep Venous Thrombosis Patient Samples: How to Perform an Easy Coagulometer Validation Process According to Available Guidelines.

Authors:  Silmara Aparecida de Lima Montalvão; Ana Paula Francisco; Bittar Letícia Queiroz da Silva; Stephany Cares Huber; Helder José Aguiari; Maria Carmem Gonçalves Lopes Fernandes; Priscila Soares Elidio; Beatriz de Moraes Martinelli; Isa Macedo Tony; Marina Pereira Colella; Joyce Maria Annichinno-Bizzachi
Journal:  Clin Appl Thromb Hemost       Date:  2020 Jan-Dec       Impact factor: 2.389

  1 in total