Literature DB >> 24639140

Long-term effectiveness of the combined minute ventilation and patient activity sensors as predictor of heart failure events in patients treated with cardiac resynchronization therapy: Results of the Clinical Evaluation of the Physiological Diagnosis Function in the PARADYM CRT device Trial (CLEPSYDRA) study.

Angelo Auricchio1, Michael R Gold, Josep Brugada, G Nölker, Siva Arunasalam, Christophe Leclercq, Pascal Defaye, Leonardo Calò, Oliver Baumann, Francisco Leyva.   

Abstract

AIMS: Monitoring early signs of clinical deterioration could allow physicians to adjust medical treatment for patients at risk of acute heart failure decompensation. To date, several strategies using different surrogate measures of clinical status emerged, but none has yet been proven to predict clinical events. We hypothesized that the Physiological Diagnostic feature, which combines data from minute ventilation and physical activity sensors, predicts heart failure events in patients implanted with cardiac resynchronization therapy with defibrillation (CRT-D) devices. METHODS AND
RESULTS: The Clinical Evaluation of the Physiological Diagnostic feature in the PARADYM CRT device (CLEPSYDRA) trial is a multicentre, prospective, non-randomized, double-blind study comprising 521 CRT-D patients with heart failure [67.4 ± 10.1 years (mean ± SD), 82% male, New York Heart Association class III/IV 85.0%/6.7%, QRS 155.3 ± 26.6 ms, left ventricular ejection fraction 25.7 ± 7.7%]. The objective of the study was the sensitivity and false positive rate of the Physiological Diagnostic algorithm to predict heart failure events within the following month. After a mean follow-up of 17.0 ± 8.7 months, 130 (25.6%) patients experienced a heart failure event. The sensitivity of the algorithm to predict an event was 34% and the false positive rate was 2.4 per patient-year.
CONCLUSION: Thirty-four per cent of heart failure events occurring within a month were predicted by the Physiological Diagnostic algorithm, and 2.4 alerts per patient per year were not followed by an heart failure event within the subsequent month.
© 2014 The Authors. European Journal of Heart Failure © 2014 European Society of Cardiology.

Entities:  

Keywords:  Activity; Cardiac resynchronization therapy; Heart failure monitoring; Minute ventilation; Predictive algorithm

Mesh:

Year:  2014        PMID: 24639140     DOI: 10.1002/ejhf.79

Source DB:  PubMed          Journal:  Eur J Heart Fail        ISSN: 1388-9842            Impact factor:   15.534


  7 in total

1.  Is There a Future for Remote Cardiac Implantable Electronic Device Management?

Authors:  Haran Burri
Journal:  Arrhythm Electrophysiol Rev       Date:  2017-08

2.  Long-term monitoring of respiratory rate in patients with heart failure: the Multiparametric Heart Failure Evaluation in Implantable Cardioverter-Defibrillator Patients (MULTITUDE-HF) study.

Authors:  Giovanni B Forleo; Luca Santini; Massimiliano Campoli; Mario Malavasi; Alberto Scaccia; Maurizio Menichelli; Umberto Riva; Filippo Lamberti; Giovanni Carreras; Serafino Orazi; Valentina Ribatti; Luigi Di Biase; Mariolina Lovecchio; Andrea Natale; Sergio Valsecchi; Francesco Romeo
Journal:  J Interv Card Electrophysiol       Date:  2015-04-28       Impact factor: 1.900

Review 3.  Clinical Utility of HeartLogic, a Multiparametric Telemonitoring System, in Heart Failure.

Authors:  Juan Carlos López-Azor; Noelia de la Torre; María Dolores García-Cosío Carmena; Pedro Caravaca Pérez; Catalina Munera; Irene MarcoClement; Rocío Cózar León; Jesús Álvarez-García; Marta Pachón; Fernando Arribas Ynsaurriaga; Rafael Salguero Bodes; Juan Francisco Delgado Jiménez; Javier de Juan Bagudá
Journal:  Card Fail Rev       Date:  2022-04-21

4.  Early Indication of Decompensated Heart Failure in Patients on Home-Telemonitoring: A Comparison of Prediction Algorithms Based on Daily Weight and Noninvasive Transthoracic Bio-impedance.

Authors:  Illapha Cuba Gyllensten; Alberto G Bonomi; Kevin M Goode; Harald Reiter; Joerg Habetha; Oliver Amft; John Gf Cleland
Journal:  JMIR Med Inform       Date:  2016-02-18

Review 5.  Accelerometer-assessed physical behavior and the association with clinical outcomes in implantable cardioverter-defibrillator recipients: A systematic review.

Authors:  Maarten Z H Kolk; Diana M Frodi; Tariq O Andersen; Joss Langford; Soeren Z Diederichsen; Jesper H Svendsen; Hanno L Tan; Reinoud E Knops; Fleur V Y Tjong
Journal:  Cardiovasc Digit Health J       Date:  2021-11-24

6.  Rationale and design of the SafeHeart study: Development and testing of a mHealth tool for the prediction of arrhythmic events and implantable cardioverter-defibrillator therapy.

Authors:  Diana M Frodi; Maarten Z H Kolk; Joss Langford; Tariq O Andersen; Reinoud E Knops; Hanno L Tan; Jesper H Svendsen; Fleur V Y Tjong; Soeren Z Diederichsen
Journal:  Cardiovasc Digit Health J       Date:  2021-10-13

7.  Haemodynamic monitoring of cardiac status using heart sounds from an implanted cardiac device.

Authors:  Pramodsingh H Thakur; Qi An; Lynne Swanson; Yi Zhang; Roy S Gardner
Journal:  ESC Heart Fail       Date:  2017-07-04
  7 in total