Francesco Fulvio Faletra1, Alamelu Ramamurthi2, Maria Cristina Dequarti3, Laura Anna Leo3, Tiziano Moccetti3, Natesa Pandian2.
Abstract
Three-dimensional (3D) transesophageal echocardiography (TEE) is subject to the same types of artifacts encountered on two-dimensional TEE. However, when displayed in a 3D format, some of the artifacts appear more "realistic," whereas others are unique to image acquisition and postprocessing. Three-dimensional TEE is increasingly used in the setting of percutaneous catheter-based interventions and ablation procedures, and 3D artifacts caused by the metallic components of catheters and devices are particularly frequent. Knowledge of these artifacts is of paramount relevance to avoid misinterpretation of 3D images. Although artifacts and pitfalls on two-dimensional echocardiography are well described and classified, a systematic description of artifacts in 3D transesophageal echocardiographic images and how they affect 3D imaging is still absent. The aim of this review is to describe the most relevant artifacts on 3D TEE, with particular emphasis on those occurring during percutaneous interventions for structural heart disease and ablation procedures.Entities:
Keywords: Artifacts; Percutaneous catheter-based interventions; Three-dimensional transesophageal echocardiography
Mesh:
Year: 2014 PMID: 24637057 DOI: 10.1016/j.echo.2014.02.003
Source DB: PubMed Journal: J Am Soc Echocardiogr ISSN: 0894-7317 Impact factor: 5.251