Effect of localizer radiograph on radiation dose associated with automatic exposure control: human cadaver and patient study.

Sarabjeet Singh1, Dean Petrovic, Ethen Jamnik, Shima Aran, Sarvenaz Pourjabbar, Maggie L Kave, Stephen E Bradley, Garry Choy, Mannudeep K Kalra.   

Abstract

PURPOSE: To evaluate the effect of localizing radiograph on computed tomography (CT) radiation dose associated with automatic exposure control with a human cadaver and patient study.
MATERIALS AND METHODS: Institutional review board approved the study with a waiver of informed consent. Two chest CT image series with fixed tube current and combined longitudinal-angular automatic exposure control (AEC) were acquired in a human cadaver (64-year-old man) after each of the 8 combinations of localizer radiographs (anteroposterior [AP], AP lateral, AP-posteroanterior [PA], lateral AP, lateral PA, PA, PA-AP, and PA lateral). Applied effective milliampere second, volume CT dose index (CTDIvol) and image noise were recorded for all 24-image series. Volume CT dose indexes were also recorded in 20 patients undergoing chest and abdominal CT after PA and PA-lateral radiographs with the use of AEC. Data were analyzed using analysis of variance and linear correlation tests.
RESULTS: With AEC, the CTDIvol fluctuates with the number and projection of localizer radiographs (P < 0.0001). Lowest CTDIvol values are seen when 2 orthogonal localizer radiographs are acquired, whereas highest values are seen when single PA or AP-PA projection localizer radiographs are acquired for planning (P < 0.0001). In 20 patients, CT scanning with AEC after acquisition of 2 orthogonal projection localizer radiographs was associated with significant reduction in radiation dose compared to PA projection radiographs alone (P < 0.0001).
CONCLUSIONS: When scanning with AEC, acquisition of 2 orthogonal localizer radiographs is associated with lower CTDIvol compared to a single localizer radiograph.

Entities:  

Mesh:

Year:  2014        PMID: 24632938     DOI: 10.1097/RCT.0000000000000035

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  7 in total

1.  Investigating the CT localizer radiograph: acquisition parameters, patient centring and their combined influence on radiation dose.

Authors:  J W Lambert; S Kumar; J S Chen; Z J Wang; R G Gould; B M Yeh
Journal:  Br J Radiol       Date:  2015-01-22       Impact factor: 3.039

2.  Organ doses from CT localizer radiographs: Development, validation, and application of a Monte Carlo estimation technique.

Authors:  Jocelyn Hoye; Shobhit Sharma; Yakun Zhang; Wanyi Fu; Francesco Ria; Anuj Kapadia; W Paul Segars; Joshua Wilson; Ehsan Samei
Journal:  Med Phys       Date:  2019-09-16       Impact factor: 4.071

3.  Reliability of body size measurements obtained at autopsy: impact on the pathologic assessment of the heart.

Authors:  Carmen A McCormack; Roberto Lo Gullo; Mannudeep K Kalra; Abner Louissaint; James R Stone
Journal:  Forensic Sci Med Pathol       Date:  2016-03-29       Impact factor: 2.007

4.  Impact of the scout view orientation on the radiation exposure and image quality in thoracic and abdominal CT.

Authors:  S Suntharalingam; A Wetter; N Guberina; J Theysohn; A Ringelstein; T Schlosser; M Forsting; K Nassenstein
Journal:  Eur Radiol       Date:  2016-03-04       Impact factor: 5.315

5.  CT dose modulation using automatic exposure control in whole-body PET/CT: effects of scout imaging direction and arm positioning.

Authors:  Yusuke Inoue; Kazunori Nagahara; Hiroko Kudo; Hiroyasu Itoh
Journal:  Am J Nucl Med Mol Imaging       Date:  2018-04-25

6.  A CONSORT-compliant prospective randomized controlled trial: radiation dose reducing in computed tomography using an additional lateral scout view combined with automatic tube current modulation: Phantom and patient study.

Authors:  Wanlin Peng; Zhenlin Li; Chunchao Xia; Yingkun Guo; Jinge Zhang; Kai Zhang; Lei Li; Fei Zhao
Journal:  Medicine (Baltimore)       Date:  2017-07       Impact factor: 1.889

7.  0° vs. 180° CT localiser: The effect of vertical off-centring, phantom positioning and tube voltage on dose optimisation in multidetector computed tomography.

Authors:  Yazan Al-Hayek; Xiaoming Zheng; Rob Davidson; Christopher Hayre; Dana Al-Mousa; Campbell Finlay; Kelly Spuur
Journal:  J Med Radiat Sci       Date:  2021-08-17
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.