Literature DB >> 24629310

Placement makes a difference: accuracy of an accelerometer in measuring step number and stair climbing.

Cornelia Lützner1, Heike Voigt1, Ingo Roeder2, Stephan Kirschner1, Jörg Lützner3.   

Abstract

UNLABELLED: Accurate and easy-to-use measurement tools are required to evaluate the effect of treatments on patient activity. Comfortable device placement and fixation are important for patient compliance. The aim of this study was the evaluation of the accuracy of an accelerometer at different placements and slow velocities.
METHODS: A total of 43 healthy volunteers were included for a literature-based treadmill protocol using five accelerometer positions; a subset of 18 volunteers performed an extended treadmill protocol with velocities between 0.1 and 2.6m/s and finally stair climbing.
RESULTS: An alternative accelerometer position at the anterolateral aspect of the middle shank did measure steps more accurately than at the manufacturer suggested position, especially during slow velocities. Participants preferred the alternative placement at the shank. The accuracy of different accelerometer positions was excellent at velocities between 1.0 and 2.2m/s. During slow velocities below 1.0m/s steps were recorded less accurately. Accepting an error of five percent, the accelerometer recorded steps accurately from 0.5m/s at the alternative placement and from 0.8m/s at the manufacturer suggested placement. Stair climbing was not recorded accurately by any accelerometer position.
CONCLUSION: For measuring step number during slow velocities, the alternative position should be favoured. Stair climbing was not recorded accurately by any tested placement.
Copyright © 2014 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Accuracy; Activities of daily living; Activity monitor; Step number

Mesh:

Year:  2014        PMID: 24629310     DOI: 10.1016/j.gaitpost.2014.01.022

Source DB:  PubMed          Journal:  Gait Posture        ISSN: 0966-6362            Impact factor:   2.840


  8 in total

1.  Constitutive Equations for Analyzing Stress Relaxation and Creep of Viscoelastic Materials Based on Standard Linear Solid Model Derived with Finite Loading Rate.

Authors:  Che-Yu Lin; Yi-Cheng Chen; Chen-Hsin Lin; Ke-Vin Chang
Journal:  Polymers (Basel)       Date:  2022-05-23       Impact factor: 4.967

2.  Patient activity after TKA depends on patient-specific parameters.

Authors:  Cornelia Lützner; Stephan Kirschner; Jörg Lützner
Journal:  Clin Orthop Relat Res       Date:  2014-07-23       Impact factor: 4.176

Review 3.  Toward Harmonized Treadmill-Based Validation of Step-Counting Wearable Technologies: A Scoping Review.

Authors:  Christopher C Moore; Aston K McCullough; Elroy J Aguiar; Scott W Ducharme; Catrine Tudor-Locke
Journal:  J Phys Act Health       Date:  2020-07-11

4.  Instrumenting gait with an accelerometer: a system and algorithm examination.

Authors:  A Godfrey; S Del Din; G Barry; J C Mathers; L Rochester
Journal:  Med Eng Phys       Date:  2015-03-04       Impact factor: 2.242

5.  Using Commercial Activity Monitors to Measure Gait in Patients with Suspected iNPH: Implications for Ambulatory Monitoring.

Authors:  Shiv Gaglani; Jessica Moore; M Ryan Haynes; Jamie B Hoffberger; Daniele Rigamonti
Journal:  Cureus       Date:  2015-11-17

6.  Validation of the ADAMO Care Watch for step counting in older adults.

Authors:  Daniele Magistro; Paolo Riccardo Brustio; Marco Ivaldi; Dale Winfield Esliger; Massimiliano Zecca; Alberto Rainoldi; Gennaro Boccia
Journal:  PLoS One       Date:  2018-02-09       Impact factor: 3.240

7.  Kinematics and temporospatial parameters during gait from inertial motion capture in adults with and without HIV: a validity and reliability study.

Authors:  Karina Berner; John Cockcroft; Quinette Louw
Journal:  Biomed Eng Online       Date:  2020-07-24       Impact factor: 2.819

8.  Independent prescribing by advanced physiotherapists for patients with low back pain in primary care: A feasibility trial with an embedded qualitative component.

Authors:  Tim Noblet; John Marriott; Amanda Hensman-Crook; Simon O'Shea; Sarah Friel; Alison Rushton
Journal:  PLoS One       Date:  2020-03-17       Impact factor: 3.240

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.