Metallic stent with high axial force as a risk factor for cholecystitis in distal malignant biliary obstruction.

Yousuke Nakai1, Hiroyuki Isayama, Kazumichi Kawakubo, Hirofumi Kogure, Tsuyoshi Hamada, Osamu Togawa, Yukiko Ito, Saburo Matsubara, Toshihiko Arizumi, Hiroshi Yagioka, Naminatsu Takahara, Rie Uchino, Suguru Mizuno, Koji Miyabayashi, Keisuke Yamamoto, Takashi Sasaki, Natsuyo Yamamoto, Kenji Hirano, Minoru Tada, Kazuhiko Koike.   

Abstract

BACKGROUND AND AIM: Tumor involvement to the orifice of cystic duct (OCD) is a risk factor for cholecystitis after self-expandable metallic stent (SEMS) placement, but its prevention is still difficult. We conducted this multicenter analysis to clarify a type of SEMS or a method to place SEMS which would decrease the incidence of cholecystitis after SEMS placement.
METHODS: The incidence of cholecystitis was studied in consecutive patients receiving SEMS for distal malignant biliary obstruction in five tertiary care centers. Multiple logistic regression analysis was performed to evaluate risk factors for cholecystitis.
RESULTS: A total of 376 patients who received SEMS placement for distal malignant biliary obstruction were analyzed. Tumor involvement to OCD was diagnosed in 25.3%. Overall incidence of cholecystitis was 6.9%. Cholecystitis was observed in 8.0% of 300 patients with covered SEMS, 16.8% of 95 patients with tumor involvement to OCD, 10.8% of 234 patients with SEMS of high axial force (AF), and 12.0% of 158 patients with SEMS length ≤ 60 mm. In the multivariate analysis, tumor involvement to OCD (odds ratio [OR] 5.40, P < 0.001), SEMSs with high AF (OR 5.33, P = 0.002), and SEMS length ≤ 60 mm (OR 3.19, P = 0.010) are risk factors. Among patients with tumor involvement to OCD, the incidence of cholecystitis in SEMS with high and low AF was 25.0% and 5.0%, respectively.
CONCLUSION: This study with an expanded cohort reconfirmed tumor involvement to OCD as a risk factor for cholecystitis after SEMS placement. SEMS with low AF might decrease cholecystitis.
© 2014 Journal of Gastroenterology and Hepatology Foundation and Wiley Publishing Asia Pty Ltd.

Entities:  

Keywords:  cholecystitis; malignant biliary obstruction; mechanical properties; self-expandable metallic stents

Mesh:

Substances:

Year:  2014        PMID: 24628054     DOI: 10.1111/jgh.12582

Source DB:  PubMed          Journal:  J Gastroenterol Hepatol        ISSN: 0815-9319            Impact factor:   4.029


  17 in total

1.  SEMS and Sensibility: Self-Expandable Metal Stents for Malignant Biliary Obstruction-Are Stent Characteristics Important?

Authors:  Evangelos Kalaitzakis
Journal:  Dig Dis Sci       Date:  2019-09       Impact factor: 3.199

2.  Endoscopic therapy for benign biliary strictures: evaluation of metal vs. plastic biliary stents.

Authors:  Joshua C Obuch; Mihir S Wagh
Journal:  Hepatobiliary Surg Nutr       Date:  2017-08       Impact factor: 7.293

Review 3.  Understanding the Mechanical forces of Self-Expandable Metal Stents in the Biliary Ducts.

Authors:  Hiroyuki Isayama; Yousuke Nakai; Tsuyoshi Hamada; Saburo Matsubara; Hirofumi Kogure; Kazuhiko Koike
Journal:  Curr Gastroenterol Rep       Date:  2016-12

Review 4.  Endoscopic Palliation of Pancreatic Cancer.

Authors:  Vishal B Gohil; Jason B Klapman
Journal:  Curr Treat Options Gastroenterol       Date:  2017-09

Review 5.  Endoscopic Management of Pancreatobiliary Malignancies.

Authors:  Dong Wook Lee; Eun Young Kim
Journal:  Dig Dis Sci       Date:  2022-02-16       Impact factor: 3.199

6.  Evaluation of mechanical properties of self-expanding metal stents for optimization of tracheal collapse in dogs.

Authors:  Ji-Hyun Kim; Jin-Young Choi; Hun-Young Yoon
Journal:  Can J Vet Res       Date:  2022-07       Impact factor: 0.897

7.  A Prospective Multicenter Study of a Fully Covered Metal Stent in Patients with Distal Malignant Biliary Obstruction: WATCH-2 Study.

Authors:  Hirofumi Kogure; Shomei Ryozawa; Iruru Maetani; Yousuke Nakai; Hiroshi Kawakami; Ichiro Yasuda; Hitoshi Mochizuki; Hirotoshi Iwano; Hiroyuki Maguchi; Mitsuhiro Kida; Kensuke Kubota; Tsuyoshi Mukai; Osamu Hasebe; Yoshinori Igarashi; Keiji Hanada; Atsushi Irisawa; Kei Ito; Takao Itoi; Hiroyuki Isayama
Journal:  Dig Dis Sci       Date:  2017-12-07       Impact factor: 3.199

8.  Feasibility of conversion of percutaneous cholecystostomy to internal transmural endoscopic ultrasound-guided gallbladder drainage.

Authors:  Tanyaporn Chantarojanasiri; Saburo Matsubara; Hiroyuki Isayama; Yousuke Nakai; Naminatsu Takahara; Suguru Mizuno; Hirofumi Kogure; Ryunosuke Hakuta; Yukiko Ito; Minoru Tada; Kazuhiko Koike
Journal:  Saudi J Gastroenterol       Date:  2017 Nov-Dec       Impact factor: 2.485

9.  Preoperative biliary drainage using a fully covered self-expandable metallic stent for pancreatic head cancer: A prospective feasibility study.

Authors:  Osamu Togawa; Hiroyuki Isayama; Hiroshi Kawakami; Yousuke Nakai; Dai Mohri; Tsuyoshi Hamada; Hirofumi Kogure; Kazumichi Kawakubo; Naoya Sakamoto; Kazuhiko Koike; Hiroto Kita
Journal:  Saudi J Gastroenterol       Date:  2018 May-Jun       Impact factor: 2.485

Review 10.  Basic Knowledge about Metal Stent Development.

Authors:  Seok Jeong
Journal:  Clin Endosc       Date:  2016-03-22
View more