| Literature DB >> 24627721 |
Lara Milevoj Kopcinovic1, Jelena Culej2.
Abstract
The pathological accumulation of serous fluids in the pleural, peritoneal and pericardial space occurs in a variety of conditions. Since patient management depends on right and timely diagnosis, biochemical analysis of extravascular body fluids is considered a valuable tool in the patient management process. The biochemical evaluation of serous fluids includes the determination of gross appearance, differentiation of transudative from exudative effusions and additional specific biochemical testing to assess the effusion etiology. This article summarized data from the most relevant literature concerning practice with special emphasis on usefulness of biochemical tests used for the investigation of pleural, peritoneal and pericardial effusions. Additionally, preanalytical issues concerning serous fluid analysis were addressed and recommendations concerning acceptable analytical practice in serous fluid analysis were presented.Entities:
Keywords: pericardial effusion; peritoneal effusion; pleural effusion
Mesh:
Substances:
Year: 2014 PMID: 24627721 PMCID: PMC3936968 DOI: 10.11613/BM.2014.014
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Figure 1.Biochemical analysis of pleural effusions.
LD – lactate dehydrogenase; URL – upper reference limit of serum LD; SEAG – serum effusion albumin gradient; NT-proBNP- N-terminal brain natriuretic peptide precursor; CHF – congestive heart failure; IMA- ischemia-modified albumin; ADA – adenosine deaminase; IFN-γ - Interferon-γ; CRP – C-reactive protein.
Light’s criteria for the differentiation of pleural effusions.
| Transudative | < 0.5 | < 0.6 | < 2/3 URL |
| Exudative | ≥ 0.5 | ≥ 0.6 | ≥ 2/3 URL |
Effusions are identified as exudative if one or more conditions are met.
LD – lactate dehydrogenase; PF – pleural fluid; URL – upper reference limit of serum LD.
Figure 2.Biochemical analysis of peritoneal effusions.
LD – lactate dehydrogenase; SAAG – serum ascites albumin gradient; SBP – spontaneous bacterial peritonitis; ADA - adenosine deaminase.
Modification of Light’s criteria for the differentiation of ascites.
| Transudative | < 0.5 | < 0.6 | < 400 |
| Exudative | ≥ 0.5 | ≥ 0.6 | ≥ 400 |
Effusions are identified as exudative if at least two conditions are met.
LD – lactate dehydrogenase.
Containers used for serous fluid biochemical analysis.
| pH analysis | heparinized syringe | sampling done in anaerobic conditions; analyze promptly using a blood gas analyzer |
| Cell count, differential cell count | EDTA-anticoagulated tube | after collection, mix well |
| Biochemical analyses | non-additive or heparin-anticoagulated containers/tubes | allow complete clotting of samples collected in non-additive containers/tubes at room temperature; bloody and turbid serous fluids may be unsuitable for analysis |
| Glucose, lactate | tubes with glycolitic inhibitors | lactate: specimen chilling during transportation to the laboratory |
EDTA - ethylenediaminetetraacetic acid.
Recommendations for acceptable practice in serous fluids analysis (10,64,75).
| • The lack of manufacturers’ claims for alternate samples obligates clinical laboratories to verify the performance of commercial assays for their possible use in serous fluid analysis. |
| • The preconditions for using a commercial assay for analysis in an alternate body fluid include that the measurement system in the routine specimen type has acceptable test characteristics, has calibrators and controls, is traceable to a reference method and has available external proficiency testing. |
| • Matrix effects, resulting from variations in serous fluid constituents’ concentrations, can alter fluid surface tension, viscosity and miscibility in a reaction mixture and potentially influence the measurements in assays intended for use in standard fluid analysis (inaccurate pipetting, inadequate mixing). This matrix effects can potentially cause erroneous results and should be given full consideration when analyzing serous fluids. |
| • Lack of reference ranges for serous fluid analysis should be compensated for by interpreting the results in conjunction with simultaneously collected serum specimens. |
| • The test report should include the measured value, type of fluid tested and a statement that accuracy might be affected by sample type, and that the obtained results should be interpreted in the clinical context. |