Hemoglobin A1c: a reliable and accurate test for diabetes care? A prospective study in Mexico.

José Gerardo González-González1, René Rodríguez-Gutiérrez2, Fernando Javier Lavalle-González1, Arnulfo González-Cantú1, Héctor Eloy Taméz-Pérez1, Gerardo González-Saldívar2, Jesús Zacarías Villarreal-Pérez1.   

Abstract

OBJECTIVE: To compare the concordance correlation coefficient for HbA1c results in an in-field experience.
MATERIALS AND METHODS: A prospective study in Monterrey, Mexico from April to August 2012 was conducted to evaluate the day-to-day clinical situation when measuring HbA1c. Blood samples from 38 consecutive patients were sent to seven local laboratories and one international reference laboratory.
RESULTS: Poor concordance was found in 4 out of 7 laboratories, moderate in 2 out of 7, and significant in just one. HbA1c values from three laboratories fluctuated more than 1% above or below the reference laboratory in more than 30% of cases, and more than 2% in 10%-20% of subjects.
CONCLUSIONS: Standardized HbA1c measurement has not occurred worldwide. Physicians should be aware of this issue and be cautious of HbA1c guidelines on diabetes diagnosis or management until proper standardization programs are implemented.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24626616     DOI: 10.21149/spm.v55i5.7245

Source DB:  PubMed          Journal:  Salud Publica Mex        ISSN: 0036-3634


  1 in total

1.  Hemoglobin A1c and risk of left atrial thrombus and spontaneous echo contrast in non-valvular atrial fibrillation patients.

Authors:  Rong-Ren Kuang; Fang-Zhou Liu; Yun-Peng Li; Wei-Dong Lin; Hua-Sheng Liang; Ai-Hua Chen
Journal:  Eur J Med Res       Date:  2017-04-21       Impact factor: 2.175

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.