Performance of a multi-axis ionization chamber array in a 1.5 T magnetic field.

K Smit1, J G M Kok, J J W Lagendijk, B W Raaymakers.   

Abstract

At the UMC Utrecht a prototype MR-linac has been installed. The system consists of an 8 MV Elekta linear accelerator and a 1.5 T Philips MRI system. This paper investigates the performance of the IC PROFILER™, a multi-axis ionization chamber array, in a 1.5 T magnetic field. The influence of the magnetic field on the IC PROFILER™ reproducibility, dose response linearity, pulse rate frequency dependence, power to electronics, panel orientation and ionization chamber shape were investigated. The linearity, reproducibility, pulse rate frequency dependence, panel orientation and ionization chamber shape are unaffected by the magnetic field. When the measurements results are normalized to the centre reference chamber, the measurements can commence unaltered. Orientation of the ionization chambers in the magnetic field is of importance, therefore caution must be taken when comparing or normalizing results from several different axes. IC PROFILER™ dose profiles were compared with film dose profiles obtained simultaneously in the MR-linac. Deviation between the film and the IC PROFILER™ data was caused by the noise in the film, indicating correct performance of the IC PROFILER™ in the transverse 1.5 T magnetic field.

Entities:  

Mesh:

Year:  2014        PMID: 24625540     DOI: 10.1088/0031-9155/59/7/1845

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  A Method to Determine the Coincidence of MRI-Guided Linac Radiation and Magnetic Isocenters.

Authors:  Kujtim Latifi; Eduardo G Moros; Geoffrey Zhang; Louis Harrison; Vladimir Feygelman
Journal:  Technol Cancer Res Treat       Date:  2019-01-01

Review 2.  Novel methodologies for dosimetry audits: Adapting to advanced radiotherapy techniques.

Authors:  Marlies Pasler; Victor Hernandez; Núria Jornet; Catharine H Clark
Journal:  Phys Imaging Radiat Oncol       Date:  2018-03-19
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.