Improvement in clinical step and shoot intensity modulated radiation therapy delivery accuracy on an integrated linear accelerator control system.

C E Agnew1, D M Irvine2, A R Hounsell3, C K McGarry3.   

Abstract

PURPOSE: The dose delivery accuracy of 30 clinical step and shoot intensity modulated radiation therapy plans was investigated using the single integrated multileaf collimator controller of the Varian Truebeam linear accelerator (linac) (Varian Medical Systems, Palo Alto, CA) and compared with the dose delivery accuracy on a previous generation Varian 2100CD C-Series linac. METHODS AND MATERIALS: Ten prostate, 10 prostate and pelvic node, and 10 head-and-neck cases were investigated in this study. Dose delivery accuracy on each linac was assessed using Farmer ionization chamber point dose measurements, 2-dimensional planar ionization chamber array measurements, and the corresponding Varian dynamic log files. Absolute point dose measurements, fluence delivery accuracy, leaf position accuracy, and the overshoot effect were assessed for each plan.
RESULTS: Absolute point dose delivery accuracy increased by 1.5% on the Truebeam compared with the 2100CD linac. No improvement in fluence delivery accuracy between the linacs, at a gamma criterion of 3%/3 mm was measured using the 2-dimensional ionization chamber array, with median (interquartile range) gamma passing rates of 98.99% (97.70%-99.72%) and 99.28% (98.26%-99.75%) for the Truebeam and 2100CD linacs, respectively. Varian log files also showed no improvement in fluence delivery between the linacs at 3%/3 mm, with median gamma passing rates of 99.97% (99.93%-99.99%) and 99.98% (99.94%-100%) for the Truebeam and 2100CD linacs, respectively. However, log files revealed improved leaf position accuracy and fluence delivery at 1%/1 mm criterion on the Truebeam (99.87%; 99.78%-99.94%) compared with the 2100CD linac (97.87%; 91.93%-99.49%). The overshoot effect, characterized on the 2100CD linac, was not observed on the Truebeam.
CONCLUSIONS: The integrated multileaf collimator controller on the Varian Truebeam improves clinical treatment delivery accuracy of step and shoot intensity modulated radiation therapy fields compared with delivery on a Varian C-series linac. Crown
Copyright © 2014. Published by Elsevier Inc. All rights reserved.

Mesh:

Year:  2013        PMID: 24621422     DOI: 10.1016/j.prro.2013.07.003

Source DB:  PubMed          Journal:  Pract Radiat Oncol        ISSN: 1879-8500


  8 in total

Review 1.  A review of recent advances in optical fibre sensors for in vivo dosimetry during radiotherapy.

Authors:  S O'Keeffe; D McCarthy; P Woulfe; M W D Grattan; A R Hounsell; D Sporea; L Mihai; I Vata; G Leen; E Lewis
Journal:  Br J Radiol       Date:  2015-03-11       Impact factor: 3.039

2.  The use of log file analysis within VMAT audits.

Authors:  Conor K McGarry; Christina E Agnew; Mohammad Hussein; Yatman Tsang; Alan R Hounsell; Catharine H Clark
Journal:  Br J Radiol       Date:  2016-04-13       Impact factor: 3.039

3.  Leakage-Penumbra effect in intensity modulated radiation therapy step-and-shoot dose delivery.

Authors:  Grigor N Grigorov; James Cl Chow
Journal:  World J Radiol       Date:  2016-01-28

4.  An EPID-based method for comprehensive verification of gantry, EPID and the MLC carriage positional accuracy in Varian linacs during arc treatments.

Authors:  Pejman Rowshanfarzad; Conor K McGarry; Michael P Barnes; Mahsheed Sabet; Martin A Ebert
Journal:  Radiat Oncol       Date:  2014-11-26       Impact factor: 3.481

5.  The Impact of Dose Rate on the Accuracy of Step-and-Shoot Intensity-modulated Radiation Therapy Quality Assurance Using Varian 2300CD.

Authors:  Christopher F Njeh; Howard W Salmon; Claire Schiller
Journal:  J Med Phys       Date:  2017 Oct-Dec

6.  Dosimetric Effect of Jaw Tracking in Volumetric-Modulated Arc Therapy.

Authors:  Sangutid Thongsawad; Chirasak Khamfongkhruea; Chirapha Tannanonta
Journal:  J Med Phys       Date:  2018 Jan-Mar

7.  An Investigation of Multileaf Collimator Performance Dependence on Gantry Angle Using Machine Log Files.

Authors:  Jeremy L Hughes; Martin A Ebert; Conor K McGarry; Christina E Agnew; Mahsheed Sabet; Pejman Rowshanfarzad
Journal:  J Med Phys       Date:  2021-12-31

8.  Correlation of phantom-based and log file patient-specific QA with complexity scores for VMAT.

Authors:  Christina E Agnew; Denise M Irvine; Conor K McGarry
Journal:  J Appl Clin Med Phys       Date:  2014-11-08       Impact factor: 2.102

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.