Evaluation of the bubble point test of a 0.22-μm membrane filter used for the sterilizing filtration of PET radiopharmaceuticals.

Kazutaka Hayashi1, Kazumasa Douhara, Genro Kashino.   

Abstract

OBJECTIVE: We developed a bubble point test kit and investigated the bubble point test of a 0.22-μm membrane filter used for the sterilizing filtration of [(18)F]FDG, [(11)C]MET and [(11)C]PIB. The bubble point test of the Millex-GS vented filter was often difficult due to air leakage from the vented portion of this filter. Therefore, to close the vented portion of this filter simply and reliably, we investigated the use of various materials.
METHODS: The bubble point test of the Millex-GS vented filter was performed by closing the vented portion of this filter with various materials, such as vinyl tape, plastic paraffin film (parafilm), urethane elastomer adhesive mat and polyethylene foam cushion tape. Gradually, the plunger inside a syringe filled with air was pushed down to increase the pressure on the pressure gauge and the bubble point test kit. Simultaneously, the pressure when a continuous stream of air bubbles that appeared out of the 0.22-μm membrane filter was measured as the product-wetted bubble point value. Then, the plunger inside a syringe filled with 10 mL of water was pushed down to wash the 0.22-μm membrane filter. As in the case in the above-mentioned method of measuring the product-wetted bubble point, the water-wetted bubble point value was measured.
RESULTS: The use of the polyethylene foam cushion tape and a double clip could easily and reliably prevent air leakage from the vented portion of the Millex-GS vented filter. In the bubble point test of [(18)F]FDG, [(11)C]MET and [(11)C]PIB, the product-wetted bubble point values were 382.7 ± 6.9 kPa, 385.4 ± 6.2 kPa and 351.6 ± 7.6 kPa, respectively. The bubble point ratio was used to determine the minimum product-wetted bubble point value. All results of the product-wetted bubble point test were beyond the minimum product-wetted bubble point value (334.4 kPa ([(18)F]FDG), 334.4 kPa ([(11)C]MET) and 310.3 kPa ([(11)C]PIB)). Then, the water-wetted bubble point values were 396.5 ± 8.3 kPa, 395.8 ± 8.3 kPa and 390.3 ± 7.6 kPa, respectively. All results of the water-wetted bubble point test were beyond the filter manufacturer's minimum bubble point specification (344.8 kPa).
CONCLUSIONS: The bubble point test technique using the bubble point test kit was practical for routine quality control tests of PET radiopharmaceuticals.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24610680     DOI: 10.1007/s12149-014-0830-0

Source DB:  PubMed          Journal:  Ann Nucl Med        ISSN: 0914-7187            Impact factor:   2.668


  2 in total

1.  Automated production of a sphingosine-1 phosphate receptor 1 (S1P1) PET radiopharmaceutical [11C]CS1P1 for human use.

Authors:  Zonghua Luo; Jiwei Gu; Robert C Dennett; Gregory G Gaehle; Joel S Perlmutter; Delphine L Chen; Tammie L S Benzinger; Zhude Tu
Journal:  Appl Radiat Isot       Date:  2019-06-20       Impact factor: 1.513

2.  GMP-compliant (68)Ga radiolabelling in a conventional small-scale radiopharmacy: a feasible approach for routine clinical use.

Authors:  Roeland Vis; Jules Lavalaye; Ewoudt Mw van de Garde
Journal:  EJNMMI Res       Date:  2015-04-24       Impact factor: 3.138

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.