Abstract
PURPOSE: We aim to evaluate the effects of multileaf collimator (MLC) leaf width (5 mm vs. 2.5 mm) on the radiosurgery planning for the treatment of spine lesions according to the modulated techniques (intensity-modulated radiotherapy [IMRT] vs. volumetric-modulated arc therapy [VMAT]) and the complexity of the target shape.Entities:
Mesh:
Year: 2014 PMID: 24606890 PMCID: PMC3995916 DOI: 10.1186/1748-717X-9-72
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Figure 1Four types of target shapes, reflecting the complexity of lesions. (a) Type 1, (b) Type 2, (c) Type 3, and (d) Type 4.
Target and spinal cord volumes
| C5 | 4.62 | 6.37 | 7.67 | 9.90 | 7.37 |
| T5 | 14.19 | 18.60 | 23.18 | 27.69 | 9.14 |
| L2 | 30.32 | 35.02 | 38.98 | 44.45 | 9.76 |
Figure 2Established treatment plans according to the four types of target shapes, two MLC leaf widths, and two types of planning techniques. (a) Type 1, (b) Type 2, (c) Type 3, and (d) Type 4 (the green line is the 18 Gy isodose line, blue line is the 15 Gy isodose line, orange line is the 12 Gy isodose line, yellow line is the 10 Gy isodose line).
Dosimetric indices and their improvement ratios according to the MLC leaf width (2.5 mm vs. 5 mm) and intensity-modulated techniques (IMRT vs. VMAT)
| Overall | TVC | 91.83 ± 11.56 | 88.10 ± 15.23 | 5.68 ± 10.07 | 0.003 |
| CI | 1.94 ± 0.53 | 2.06 ± 0.81 | −2.42 ± 12.25 | 0.110 | |
| GI | 9.99 ± 2.13 | 10.89 ± 2.81 | −6.25 ± 18.84 | 0.023 | |
| IMRT | TVC | 88.40 ± 15.62 | 83.55 ± 20.24 | 8.38 ± 13.66 | 0.042 |
| CI | 2.03 ± 0.67 | 2.24 ± 1.06 | −4.86 ± 13.00 | 0.119 | |
| GI | 9.30 ± 2.06 | 10.98 ± 3.34 | −13.79 ± 7.38 | 0.003 | |
| VMAT | TVC | 95.26 ± 3.12 | 92.65 ± 5.48 | 2.97 ± 3.10 | 0.005 |
| CI | 1.85 ± 0.34 | 1.88 ± 0.41 | 0.02 ± 11.48 | 0.689 | |
| GI | 10.68 ± 2.04 | 10.80 ± 2.30 | 1.27 ± 23.74 | 0.871 |
Figure 3Comparison of TVC between 2.5-mm and 5-mm MLC in IMRT and VMAT plans.
Comparison of V between IMRT and VMAT plans according to the MLC leaf width and the types of target
| MLC | 2.5-mm MLC | 455.84 ± 316.45 | 466.10 ± 288.69 | 0.875 |
| 5.0-mm MLC | 465.76 ± 315.02 | 466.78 ± 276.48 | 0.875 | |
| Target shapes | Type 1 | 315.33 ± 216.79 | 380.36 ± 257.99 | 0.075 |
| Type 2 | 439.84 ± 287.24 | 470.90 ± 290.22 | 0.173 | |
| Type 3 | 551.47 ± 368.23 | 514.65 ± 305.96 | 0.249 | |
| Type 4 | 536.55 ± 361.91 | 499.86 ± 296.27 | 0.249 | |
| Overall | 460.80 ± 308.84 | 466.44 ± 273.95 | 1.000 |
Improvement ratios of the dosimetric indices according to the types of target in IMRT and VMAT
| Overall | TVC | 0.42 ± 1.01 | 2.39 ± 5.24 | 2.74 ± 2.80 | 17.16 ± 14.68 | 0.006 |
| CI | −0.41 ± 9.34 | −5.58 ± 6.61 | 3.77 ± 9.00 | −7.46 ± 19.63 | 0.396 | |
| GI | 8.89 ± 33.04 | −10.61 ± 4.40 | −7.85 ± 5.31 | −15.46 ± 9.52 | 0.122 | |
| IMRT | TVC | −0.16 ± 0.36 | 3.62 ± 7.97 | 2.10 ± 2.32 | 27.99 ± 13.35 | 0.007 |
| CI | −4.97 ± 8.94 | −6.83 ± 4.75 | 7.16 ± 9.91 | −14.81 ± 18.92 | 0.231 | |
| GI | −8.85 ± 4.59 | −14.30 ± 2.43 | −9.37 ± 5.68 | −22.62 ± 7.57 | 0.047 | |
| VMAT | TVC | 1.01 ± 1.18 | 1.15 ± 0.78 | 3.37 ± 3.61 | 6.34 ± 3.11 | 0.094 |
| CI | 4.14 ± 8.72 | −4.33 ± 9.06 | −0.39 ± 8.35 | −0.11 ± 21.06 | 0.883 | |
| GI | 26.64 ± 41.98 | −6.92 ± 1.32 | −6.33 ± 5.59 | −8.29 ± 3.92 | 0.208 | |
Figure 4Improvement ratio of TVC according to the types of target.