Comparison of visual quality between aspheric and spherical IOLs.

Ramazan Yagci1, Feyza Uzun, Semra Acer, Ibrahim F Hepsen.   

Abstract

PURPOSE: To determine if aspheric intraocular lens (IOL) implantation produces the same degree of postoperative ocular aberration and contrast sensitivity as spherical IOL implantation.
METHODS: In this randomized prospective comparative study, 60 eyes of 30 cataract surgery patients were randomly assigned to receive a spherical IOL (Rayner 620H) in one eye and an aspheric IOL (Rayner 920H) in the contralateral eye. All patients were examined at 1 month postoperatively. Primary outcomes of contrast sensitivity and ocular wavefront higher order aberrations (HOAs) were assessed.
RESULTS: Aspheric IOLs (median total HOAs 0.26 root mean square [RMS]; range 0.13-0.82 RMS) produced significantly lower total HOAs than spherical IOLs (median total HOAs 0.34 RMS; range 0.18-1.08 RMS; p<0.05). Contrast sensitivity was significantly better with aspheric IOLs (median contrast sensitivity 1.8 log units; range 1.35-1.8 log units) than with spherical IOLs (median contrast sensitivity 1.65 log units; range 1.35-1.8 log units; p<0.05).
CONCLUSIONS: When compared with a structurally (platform and material) similar spherical IOL (Rayner 620H), aspheric IOLs (Rayner 920H) appear to significantly reduce HOAs and yield better levels of contrast sensitivity under photopic conditions.

Entities:  

Mesh:

Year:  2014        PMID: 24604606     DOI: 10.5301/ejo.5000452

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  2 in total

1.  Comparison of Ocular Monochromatic Higher-order Aberrations in Normal Refractive Surgery Candidates of Arab and South Asian Origin.

Authors:  Gaurav Prakash; Dhruv Srivastava; Sounak Choudhuri; Ruthchel Bacero
Journal:  Middle East Afr J Ophthalmol       Date:  2016 Jan-Mar

2.  Effect of Spherical Aberration on the Optical Quality after Implantation of Two Different Aspherical Intraocular Lenses.

Authors:  Michael Lasta; Kata Miháltz; Illés Kovács; Pia Veronika Vécsei-Marlovits
Journal:  J Ophthalmol       Date:  2017-08-16       Impact factor: 1.909

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.