BACKGROUND: The present study was conducted to assess the efficiency of using TK SLC-L (Salubris, Inc.) for the primary isolation of mycobacteria from clinical samples by comparing it to the MGIT detection system (Becton Dickinson Diagnostic Instrument Systems). Although TK SLC, a biphasic medium, has been evaluated previously, this is the first study to evaluate TK SLC-L, a liquid medium.Entities:
Mesh:
Substances:
Year: 2014 PMID: 24597726 PMCID: PMC3946173 DOI: 10.1186/1471-2334-14-130
Source DB: PubMed Journal: BMC Infect Dis ISSN: 1471-2334 Impact factor: 3.090
Comparison of microscopy, MGIT, and TK SLC-L for the detection of mycobacteria from clinical samples
| Total positive results among the 146 samples | 11 | 7.5 | 35 | 24.0 | 34 | 23.3 |
| Among 46 culture-positive samples | 11 | 23.9 | 35 | 76.1 | 34 | 73.9 |
| Contamination | NA | NA | 2 | 1.3 | 20 | 13.7 |
| Min/max duration for growth detection (days) | NA | NA | 5.2/34.8 | 2.0/34.1 | ||
| Mean time to growth detection (days) | NA | NA | 18.3 | 13.1 | ||
| Median time to growth detection (days) | NA | NA | 15.1 | 7.7 | ||
| Sensitivity* (%) | 23.9 | 76.1 | 73.9 | |||
| Specificity* (%) | 100 | 100 | 100 | |||
| Positive predictive value* (%) | 100 | 100 | 100 | |||
| Negative predictive value* (%) | 74.1 | 89.3 | 88.0 | |||
| Accuracy* (%) | 76.0 | 92.5 | 91.8 | |||
| Correlation** (%) | 42.1 | 80.6 | 53.0 | |||
*Sensitivity, specificity, positive predictive, negative predictive, and accuracy values were calculated according to total mycobacterial isolation.
**Pearson’s correlation.
NA, Not applicable.
Comparison of MYCOLOR TK and MGIT system
| Working principle | Colorimetric | Fluorometric |
| Sample capacity | 336 to 2016 (modular) | 960 |
| Reading intervals | Adjustable (default set every two hours) | Once in a hour |
| Approximated test cost** | 1.2-1.9 USD | 4,5-7 USD |
| Approximated susceptibility and typing cost** | 7.5-10.0 USD | 18-22,5 USD |
| Shelf life | 6-9 months | More than 1 year |
| Storage temperature for culture media | 2-8°C | Room temperature |
| Storage temperature for supplements | 2-8 C (Supplements included in TK media) | 2-8 C |
| Online access from internet | Via epicenter | Via epicenter |
| Yes | Via extra test | |
| Mycobacteria contamination differentiation | Via colorimetric or visual | Via turbidity or microscopy |
| Growth curves; auto-evaluation | Yes | Yes |
| Statistics | Yes | Via epicenter |
| Alternative visual evaluation without instrument | Yes | No |
*Table was arranged according to manufacturer’s instructions.
**Approximated test cost, susceptibility and typing cost vary according to annual consumption amounts.