Literature DB >> 24597677

Validation of Sysmex XT-2000iV generated quantitative bone marrow differential counts in untreated Wistar rats.

Kay A Criswell1, Jeffrey H Bock, Samantha E Wildeboer, Kjell Johnson, Richard P Giovanelli.   

Abstract

BACKGROUND: Preclinical drug trials frequently require assessment of bone marrow toxicity in animals to evaluate hematopoietic safety. Since the gold standard, cytologic evaluation, is time consuming and requires highly trained individuals, automated methods remain intriguing.
OBJECTIVE: The Sysmex XT-2000iV hematology analyzer allows user-developed customizable gating. This study was conducted to validate the gating of bone marrow cell populations in Sysmex cytograms from untreated rats.
METHODS: B- and T-lymphocytes and myeloid cells were experimentally depleted from Charles River Wistar Han IGS (CRL: WI [Han]) rat whole bone marrow suspension using a magnetic cell sorting (MACS) method. The positively and negatively selected populations were used to verify select gates within the Sysmex cytogram. Intra- and inter-animal precision, comparability between right and left femur, as well as agreement with microscopic myelograms based on 500 counted cells, were determined.
RESULTS: Intra-sample precision and right-to-left femur comparability confirmed that gating was reproducible and stable. In 50 tested rats, myeloid to erythroid ratios (M:E) were 1.32 ± 0.33 in males and 1.38 ± 0.29 in females by Sysmex compared to 1.36 ± 0.32 in males and 1.42 ± 0.32 in females by microscopic evaluations. Bland-Altman differences between methods was ≤ ± 0.35 units for M:E, ≤ 5.4% for maturing myeloid cells, ≤ 3.4% for proliferating myeloid cells, ≤ 6.0% for maturing myeloid cells, ≤ 3.4% for proliferating myeloid cells, and ≤ 4.1% for lymphocytes.
CONCLUSIONS: In untreated control Charles River Wistar Han IGS (CRL: WI [Han]) rats, the Sysmex XT-2000iV produced an automated M:E and 5-part differential count equivalent to microscopic differential counts.
© 2014 American Society for Veterinary Clinical Pathology and European Society for Veterinary Clinical Pathology.

Entities:  

Keywords:  Flow cytometry; hematotoxicity; myelotoxicity; preclinical development; preclinical safety

Mesh:

Year:  2014        PMID: 24597677     DOI: 10.1111/vcp.12132

Source DB:  PubMed          Journal:  Vet Clin Pathol        ISSN: 0275-6382            Impact factor:   1.180


  2 in total

1.  Reliable assessment of bone marrow and bone marrow concentrates using automated hematology analyzer.

Authors:  Venkata P Mantripragada; Nicolas S Piuzzi; Jaiben George; Wesley Bova; Mitchell Ng; Cynthia Boehm; George F Muschler
Journal:  Regen Med       Date:  2019-07-19       Impact factor: 3.806

2.  Hematologic and serum biochemical reference intervals of the Oriental white stork (Ciconia boyciana) and the application of an automatic hematologic analyzer.

Authors:  Jae-Ik Han; Hye-Jin Jang; Ki-Jeong Na
Journal:  J Vet Sci       Date:  2016-09-30       Impact factor: 1.672

  2 in total