Literature DB >> 24596406

The precision and accuracy of Nellcor and Masimo oximeters at low oxygen saturations (70%) in newborn lambs.

J A Dawson, P Bastrenta, F Cavigioli, M Thio, T Ong, M L Siew, S B Hooper, P G Davis.   

Abstract

BACKGROUND: Administration of oxygen in the delivery room is informed by oxygen saturation (SpO₂). An oxygen saturation range of 60%-70% is the threshold for administering oxygen in the first minutes after birth. The accuracy of newer generation oximeters to measure SpO₂ has not been compared against the 'gold standard', direct arterial blood oxygen saturation (SaO₂) when SaO₂ is low. The aim of this study was to determine the accuracy and precision of Nellcor and Masimo oximeters to measure SpO₂ when SaO₂ <70%.
METHOD: Prospective observational study in ventilated anaesthetised newborn lambs with an indwelling carotid artery catheter. Ventilation was adjusted to achieve hypoxaemia. Nellcor (Oxi-Max 600 with Max-N sensor) and Masimo (Rad 4 with low noise optical probe (LNOP) sensor) sensors were applied to the right forelimb (preductal). An arterial blood sample was collected at 1-5 min intervals when the animal was hypoxic. The displayed SpO₂ was recorded. We used Bland-Altman analysis to determine precision and accuracy of each oximeter when SaO₂ <70%.
RESULTS: 17 lambs were studied, 165 measurements were obtained, 123 were SaO₂ <70%. The mean difference (±1.96 SD) Nellcor SpO₂-SaO₂ when SaO₂ <70% was 17% (-12% to 46%). The mean difference (±1.96 SD) Masimo SpO₂-SaO₂ when SaO2 <70% was 13% (-19% to 45%).
CONCLUSIONS: At SaO₂<70%, both monitors overestimated oxygen saturation (SpO₂) compared with the gold standard. Both oximeters were equally inaccurate when SaO₂ was low.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24596406     DOI: 10.1136/archdischild-2013-305091

Source DB:  PubMed          Journal:  Arch Dis Child Fetal Neonatal Ed        ISSN: 1359-2998            Impact factor:   5.747


  4 in total

Review 1.  Intermittent Hypoxemia in Preterm Infants.

Authors:  Juliann M Di Fiore; Peter M MacFarlane; Richard J Martin
Journal:  Clin Perinatol       Date:  2019-06-15       Impact factor: 3.430

Review 2.  Intermittent hypoxemia and oxidative stress in preterm infants.

Authors:  Juliann M Di Fiore; Maximo Vento
Journal:  Respir Physiol Neurobiol       Date:  2019-05-14       Impact factor: 1.931

Review 3.  Standards for preclinical research and publications in developmental anaesthetic neurotoxicity: expert opinion statement from the SmartTots preclinical working group.

Authors:  Gregory A Chinn; Matthew L Pearn; Laszlo Vutskits; Cyrus D Mintz; Andreas W Loepke; Jennifer J Lee; Jerri Chen; Zeljko J Bosnjak; Ansgar M Brambrink; Vesna Jevtovic-Todorovic; Lena S Sun; Jeffrey W Sall
Journal:  Br J Anaesth       Date:  2020-03-04       Impact factor: 9.166

4.  Monitoring Oxygenation and Gas Exchange in Neonatal Intensive Care Units: Current Practice in the Netherlands.

Authors:  Ratna N G B Tan; Estelle E M Mulder; Enrico Lopriore; Arjan B Te Pas
Journal:  Front Pediatr       Date:  2015-11-03       Impact factor: 3.418

  4 in total