Better inflation time of stent balloon for second-generation drug-eluting stent expansion and apposition: an optical coherence tomography study.

Soe Hee Ann1, Jae-Wook Chung, Cai DE Jin, Jun Ho Lee, Jong Min Kim, Scot Garg, Eun-Seok Shin.   

Abstract

OBJECTIVES: We tried to determine the effect of stent balloon inflation time on stent expansion and apposition using optical coherence tomography.
BACKGROUND: Second-generation drug-eluting stents (DES) have thin struts; however, inflation times for optimal stent expansion and apposition are unknown in vivo.
METHODS: Subjects included 17 patients (18 de novo coronary artery lesions), in whom Resolute Integrity → (n = 9) and Xience Prime → (n = 9) DES were deployed. All stents were inflated 3 times to the nominal inflation pressure (8.9 ± 0.6 atm) using the stent delivery balloon. The first inflation continued until the stent was angiographically fully expanded; the other 2 lasted 15 and 30 seconds, respectively.
RESULTS: After the first, second, and third inflation of stent balloon, stent area (5.94 ± 1.7, 6.69 ± 1.8, 7.05 ± 1.8 mm(2) , P < 0.001) and stent volume (146.94 ± 59.40, 166.78 ± 69.55, 177.25 ± 69.19 mm(3) , P < 0.001) increased significantly. The number of malapposed struts (18.0 ± 17.0, 7.9 ± 10.2, 7.4 ± 10.8, P < 0.001) and the mean depth of malapposed struts (188.9 ± 75.6, 120.3 ± 101.4, 95.4 ± 86.8 µm, P < 0.001) decreased. Malapposed stent area (0.62 ± 0.32, 0.52 ± 0.21 mm(2) , P < 0.05) and the malapposed stent volume (15.03 ± 7.78, 12.64 ± 5.16 mm(3) , P < 0.05) decreased significantly following the second inflation; the third inflation gave no additional benefits to these parameters. There was no adverse clinical outcome after each stent balloon inflation.
CONCLUSIONS: Additional 15 seconds of inflation after the angiographically full expansion of the stent balloon allows better stent expansion and apposition even though the inflation pressure is nominal pressure.
© 2014, Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2014        PMID: 24593834     DOI: 10.1111/joic.12096

Source DB:  PubMed          Journal:  J Interv Cardiol        ISSN: 0896-4327            Impact factor:   2.279


  2 in total

1.  Impact of the Balloon Inflation Time and Pattern on the Coronary Stent Expansion.

Authors:  Jarosław Skowroński; Rafał Wolny; Jan Jastrzębski; Paweł Tyczyński; Karol Szlazak; Jerzy Pręgowski; Gary S Mintz; Karolina Liżewska; Wojciech Świeszkowski; Zbigniew Chmielak; Adam Witkowski
Journal:  J Interv Cardiol       Date:  2019-03-21       Impact factor: 2.279

2.  A randomized controlled clinical trial of prolonged balloon inflation during stent deployment strategy in primary percutaneous coronary intervention for ST-segment elevation myocardial infarction: a pilot study.

Authors:  Min Ma; Ling Wang; Kai-Yue Diao; Shi-Chu Liang; Ye Zhu; Hua Wang; Mian Wang; Li Zhang; Zhi-Gang Yang; Yong He
Journal:  BMC Cardiovasc Disord       Date:  2022-02-04       Impact factor: 2.298

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.