A dosimetric characterization of a novel linear accelerator collimator.

C M Thompson1, S J Weston1, V C Cosgrove1, D I Thwaites2.   

Abstract

PURPOSE: The aim of this work is to characterize a new linear accelerator collimator which contains a single pair of sculpted diaphragms mounted orthogonally to a 160 leaf multileaf collimator (MLC). The diaphragms have "thick" regions providing full attenuation and "thin" regions where attenuation is provided by both the leaves and the diaphragm. The leaves are mounted on a dynamic leaf guide allowing rapid leaf motion and leaf travel over 350 mm.
METHODS: Dosimetric characterization, including assessment of leaf transmission, leaf tip transmission, penumbral width, was performed in a plotting tank. Head scatter factor was measured using a mini-phantom and the effect of leaf guide position on output was assessed using a water phantom. The tongue and groove effect was assessed using multiple exposures on radiochromic film. Leaf reproducibility was assessed from portal images of multiple abutting fields.
RESULTS: The maximum transmission through the multileaf collimator is 0.44% at 6 MV and 0.52% at 10 MV. This reduced to 0.22% and 0.27%, respectively, when the beam passes through the dynamic leaf guide in addition to the MLC. The maximum transmission through the thick part of the diaphragm is 0.32% and 0.36% at 6 and 10 MV. The combination of leaf and diaphragm transmission ranges from 0.08% to 0.010% at 6 MV and 0.10% to 0.14% depending on whether the shielding is through the thick or thin part of the diaphragm. The off-axis intertip transmission for a zero leaf gap is 2.2% at 6 and 10 MV. The leaf tip penumbra for a 100 × 100 mm field ranges from 5.4 to 4.3 mm at 6 and 10 MV across the full range of leaf motion when measured in the AB direction, which reduces to 4.0-3.4 mm at 6 MV and 4.5-3.8 mm at 10 MV when measured in the GT direction. For a 50 × 50 mm field, the diaphragm penumbra ranges from 4.3 to 3.7 mm at 6 MV and 4.5 to 4.1 mm at 10 MV in the AB direction and 3.7 to 3.2 mm at 6 MV and 4.2 to 3.7 mm when measured in the GT direction. The tongue and groove effect observed from exposure of a radiochromic film to two abutting fields is an underdose of 25%. The head scatter factor at both 6 and 10 MV is similar to that from the MLCi2 collimator to within 0.8%. The uncertainty in the leaf position reproducibility is 0.05 mm (2σ).
CONCLUSIONS: The Agility collimator is a low leakage, high definition collimator where both the MLC and the sculpted diaphragm have been optimized for dynamic treatments.

Entities:  

Mesh:

Year:  2014        PMID: 24593717     DOI: 10.1118/1.4866228

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  13 in total

1.  Dynamic positioning accuracy of a novel multileaf collimator for volumetric modulated arc therapy.

Authors:  Yuji Nakaguchi; Takeshi Ono; Ryota Onizuka; Masato Maruyama; Yoshinobu Shimohigashi; Yudai Kai
Journal:  Radiol Phys Technol       Date:  2015-11-26

2.  Investigation of two linear accelerator head designs for treating brain metastases with hypofractionated volumetric-modulated arc radiotherapy.

Authors:  Mark Ruschin; Arjun Sahgal; Sara Iradji; Hany Soliman; Claudia Leavens; Young Lee
Journal:  Br J Radiol       Date:  2016-04-13       Impact factor: 3.039

3.  Dosimetry and evaluating the effect of treatment parameters on the leakage of multi leaf collimators in ONCOR linear accelerators.

Authors:  Keyvan Jabbari; Muhaddeseh Akbari; Mohamad Bagher Tavakoli; Alireza Amouheidari
Journal:  Adv Biomed Res       Date:  2016-12-27

4.  Commissioning an Elekta Versa HD linear accelerator.

Authors:  Ganesh Narayanasamy; Daniel Saenz; Wilbert Cruz; Chul S Ha; Niko Papanikolaou; Sotirios Stathakis
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

5.  Dosimetric validation of Monaco treatment planning system on an Elekta VersaHD linear accelerator.

Authors:  Ganesh Narayanasamy; Daniel L Saenz; Dewayne Defoor; Niko Papanikolaou; Sotirios Stathakis
Journal:  J Appl Clin Med Phys       Date:  2017-09-25       Impact factor: 2.102

6.  Pinnacle3 modeling and end-to-end dosimetric testing of a Versa HD linear accelerator with the Agility head and flattening filter-free modes.

Authors:  Daniel L Saenz; Ganesh Narayanasamy; Wilbert Cruz; Nikos Papanikolaou; Sotirios Stathakis
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

7.  Efficient independent planar dose calculation for FFF IMRT QA with a bivariate Gaussian source model.

Authors:  Feifei Li; Ji-Yeon Park; Brendan Barraclough; Bo Lu; Jonathan Li; Chihray Liu; Guanghua Yan
Journal:  J Appl Clin Med Phys       Date:  2017-02-28       Impact factor: 2.102

8.  Modeling the Agility MLC in the Monaco treatment planning system.

Authors:  Michael Snyder; Robert Halford; Cory Knill; Jeffrey N Adams; Todd Bossenberger; Adrian Nalichowski; Ahmad Hammoud; Jay Burmeister
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

9.  Intensity modulated arc therapy implementation in a three phase adaptive (18)F-FDG-PET voxel intensity-based planning strategy for head-and-neck cancer.

Authors:  Dieter Berwouts; Luiza Ana Maria Olteanu; Bruno Speleers; Frédéric Duprez; Indira Madani; Tom Vercauteren; Wilfried De Neve; Werner De Gersem
Journal:  Radiat Oncol       Date:  2016-04-02       Impact factor: 3.481

10.  Lateral Penumbra Modelling Based Leaf End Shape Optimization for Multileaf Collimator in Radiotherapy.

Authors:  Dong Zhou; Hui Zhang; Peiqing Ye
Journal:  Comput Math Methods Med       Date:  2016-01-19       Impact factor: 2.238

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.