[Hypoxic gas mixture delivery due to a defective vaporiser manifold: case report, review of the literature and suggested emergency management algorithm].

T Berlet1.   

Abstract

A case of delivery of a hypoxic gas mixture to a patient during total intravenous anesthesia is described. A progressive fall in inspiratory oxygen concentration followed by a drop in oxygen saturation below 90 % occurred during the advanced stages of a hitherto uneventful general anesthesia of a female patient undergoing anterior cervical fusion surgery. A malfunctioning defective rubber seal of a vaporizer manifold was identified as the cause of the gas leak. The leak had not been detected during the preanesthesia leak test. The problem of hypoxic gas mixtures and uncommon leaks in modern anesthesia equipment is discussed. The importance of locating a leak in the high or low pressure circuits is explained. An algorithm for the management of an unexpected decrease of inspiratory oxygen concentration or any other manifestation of a gas leak along with a systematic approach to locating the source of a gas leak is presented.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24584839     DOI: 10.1007/s00101-014-2306-y

Source DB:  PubMed          Journal:  Anaesthesist        ISSN: 0003-2417            Impact factor:   1.041


  7 in total

1.  Acute failure of oxygen delivery.

Authors:  B C Ong; J bin Katjo; B L Tan; C C Lee; Y W Chan
Journal:  Anesthesiology       Date:  2001-10       Impact factor: 7.892

2.  Aestiva ventilation mode selector switch failures.

Authors:  Dietrich Gravenstein; Harshdeep Wilkhu; Edwin B Liem; Stuart Tilman; Samsun Lampotang
Journal:  Anesth Analg       Date:  2007-04       Impact factor: 5.108

3.  Unusual cause of leak in Datex Aisys.

Authors:  Prashant Kummar; Grace Korula; Sathish Kumar; P A Saravanan
Journal:  Anesth Analg       Date:  2009-10       Impact factor: 5.108

4.  An uncommon leak in the anesthesia breathing circuit.

Authors:  H R Bromley; S Tuorinsky
Journal:  Anesth Analg       Date:  1997-09       Impact factor: 5.108

5.  Hypoxic gas mixture delivery due to malfunctioning inlet port of a Select-a-Tec vaporiser manifold.

Authors:  J Riendl
Journal:  Can J Anaesth       Date:  1987-07       Impact factor: 5.063

6.  Patient injuries from anesthesia gas delivery equipment: a closed claims update.

Authors:  Sonya P Mehta; James B Eisenkraft; Karen L Posner; Karen B Domino
Journal:  Anesthesiology       Date:  2013-10       Impact factor: 7.892

7.  Delivery of an hypoxic gas mixture due to a defective rubber seal of a flowmeter control tube.

Authors:  H Hay
Journal:  Eur J Anaesthesiol       Date:  2000-07       Impact factor: 4.330

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.